Kong Shuying,Zhou Qiang,Tuo Fei,et al.Comparative analysis of statistical measures to assess the capability of the γ-ray spectroscopy of radionuclides for interlaboratory comparison[J].Chinese Journal of Radiological Medicine and Protection,2025,45(6):573-578
Comparative analysis of statistical measures to assess the capability of the γ-ray spectroscopy of radionuclides for interlaboratory comparison
Received:October 17, 2024  
DOI:10.3760/cma.j.cn112271-20241017-00402
KeyWords:Interlaboratory comparison  γ-ray spectrum  Capability verification  Statistical measure for assessment
FundProject:
Author NameAffiliationE-mail
Kong Shuying Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Zhou Qiang Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Tuo Fei Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China tuofei@nirp.chinacdc.cn 
Yang Baolu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Li Zeshu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Qin Weihao Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Hits: 137
Download times: 45
Abstract::
      Objective To explore the applicability of critical statistical measures Z, Z', Zeta (ζ), and En value in comparative capability assessments by analyzing the data from nationwide radiological health technology institutions from 2020 to 2023, aiming to provide a reference for further improving China's comparative assessment method for verifying the γ-ray spectroscopy capability of radionuclides. Methods Based on the nationwide analysis and comparison of γ-ray spectroscopy of radionuclides conducted annually by the National Institute for Radiological Protection, China CDC, the deviations between laboratory result and reference values were determined using the above-mentioned four statistical measures. Finally, the qualification rates of the γ-ray spectroscopy result from relevant institutions were assessed and analyzed. Results The result indicate that the qualification rates of the Z value exceeded 85%, which might be inaccurate in some cases since it ignored the limitation of uncertainty. In contrast, the Z' value yielded more scientific and representative result because it considered the uncertainty of reference values, and the inter-laboratory comparison qualification rate is above 95%. In the case of low uncertainty of the γ-ray spectroscopy result, the Zeta (ζ) or En ratio exhibited relatively high sensitivity and, accordingly, relatively low qualification rates, indicating that this ratio may not comprehensively reflect the interlaboratory comparison result. Conclusions The Z, Z', Zeta (ζ), and En values demonstrate their own advantages and limitations in assessing the capability of the γ-ray spectroscopy of radionuclides. For comparison projects involving multiple laboratories and complex measurement conditions, such as the nationwide verification of the capabilities of the γ-ray spectroscopy of radionuclides, it is feasible to combine the Z' value with the Zeta (ζ) or En ratio, with the former and the latter used to determine the qualified and excellent capability result, respectively. This enables the comprehensive assessment of the measurement and analysis capabilities of the γ-ray spectroscopy laboratories of radiological health technology institutions in China.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12036275  On-line:0

v
Scan QR Code
&et=CA2056D32C3CC3425C4A7B1BF7EF20B11E5F0EAE025435ABE92BB45FF166B0670687EDBB97EDE49F3B7554E6FEE11A2F74CEF18BA01090B76AA3F541AFE6B7C592851A2E9E6FE73874C9C7F200191718E8D147A9003F679DF171B5270FD290868B5891187B29B796821F00D4979471D4E5169EB64A66A1508A1F1DAFCE2B6752D4897C5F2FC0123724038ABB3042CBE109CEE8D78A349939&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=E6A891D2134F30C0&aid=&vid=&iid=B31275AF3241DB2D&sid=03EE8EDD44A3D4BE&eid=C2053D4E59904B8A&fileno=20250612&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="E6A891D2134F30C0"; var my_aid="";