Zeng Hongwei,Zhang Yu,E Xiangyu,et al.Dosimetric evaluation of different aperture shape controller parameters based on the Halcyon 3.0 accelerator in the hippocampal avoidance-whole brain radiotherapy[J].Chinese Journal of Radiological Medicine and Protection,2023,43(9):682-688
Dosimetric evaluation of different aperture shape controller parameters based on the Halcyon 3.0 accelerator in the hippocampal avoidance-whole brain radiotherapy
Received:February 21, 2023  
DOI:10.3760/cma.j.issn.112271-20230221-00047
KeyWords:Halcyon accelerator  Hippocampal avoidance  Whole brain radiotherapy  Shape control parameter  Dosimetry
FundProject:上海中医药大学附属曙光医院中西医结合研究专项(SGZXY-202201)
Author NameAffiliationE-mail
Zeng Hongwei Department of Radiotherapy, Shuguang Hospital Affiliated to Shanghai University of Traditional Chinese Medicine, Shanghai 201203, China  
Zhang Yu Department of Radiotherapy, Shuguang Hospital Affiliated to Shanghai University of Traditional Chinese Medicine, Shanghai 201203, China  
E Xiangyu Department of Radiotherapy, Shuguang Hospital Affiliated to Shanghai University of Traditional Chinese Medicine, Shanghai 201203, China  
Feng Yue Department of Radiotherapy, Shuguang Hospital Affiliated to Shanghai University of Traditional Chinese Medicine, Shanghai 201203, China  
Yu Jingping Department of Radiotherapy, Shuguang Hospital Affiliated to Shanghai University of Traditional Chinese Medicine, Shanghai 201203, China Yujingping700420@sina.com 
Hits: 3446
Download times: 1019
Abstract::
      Objective To evaluate the dosimetric effects of different aperture shape controller (ASC) parameters based on the Halcyon 3.0 accelerator in the hippocampal avoidance-whole brain radiotherapy (HA-WBRT) plans.Methods This study enrolled 13 patients treated with WBRT using a Halcyon 3.0 accelerator at the Department of Radiotherapy of the Shuguang Hospital Affiliated to Shanghai University of Traditional Chinese Medicine. According to different aperture shape controller (ASC) parameters, the HA-WBRT plans were divided into six groups, namely non, very low, low, moderate, high, and very high ASC settings, marked by AO, AVL, AL, AM, AH, and AVH, respectively. This study presented a statistical evaluation of the effects of different ASC parameters on conformity index (CI), homogeneity index (HI), radiation doses to the hippocampus and other organs at risk, and the plans' execution efficiency. Furthermore, using two dose verification tools, namely ArcCHECK and Portal Dosimetry, this study delivered a comprehensive analysis of the differences in beam delivery precision of the plans with different ASC parameters.Results The six groups all met the requirements for clinical treatment. They had similar CI and HI values, with no statistically significant differences (P > 0.05). The AVH group showed better protection effects on organs at risk. Compared to the control group, the AVL group showed lower Dmax of brainstems but higher Dmax of chiasma opticum (F=6.26, 8.04, P < 0.05). Compared to the control group, the AH group showed lower Dmax of eyeballs but higher Dmax of optic nerves (F=2.04, 1.37, P < 0.05). In contrast, the AVH group exhibited lower Dmax of brainstems, eyeballs, and lens than the control group (F=6.26, 2.04, 2.02, P < 0.05). No statistically significant differences were observed in dosimetric indices of other organs at risk (P > 0.05). As verified using ArcCHECK and Portal Dosimetry, the γ passing rates of the six groups were over 98% at 2%/2 mm and 100% at 3%/3 mm. The overall γ passing rates verified using ArcCHECK were lower than those verified using Portal Dosimetry. The maximum difference in the monitor unit among the six groups was less than 15, and these groups did not show significant differences in terms of execution efficiency.Conclusions The HA-WBRT plans based on the Halcyon 3.0 accelerator can meet the requirements for clinical treatment. Different ASC parameters can significantly optimize the dosimetric parameters. Among them, the AVH parameters can highly reduce the radiation dose to organs at risk. Furthermore, different ASC parameters show insignificant effects on beam delivery precision and plan execution efficiency, meeting the verification standards for clinical therapeutic doses.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12034924  On-line:0

v
Scan QR Code
&et=8400690B225D0A44B78A0B7A1006E4B55F82552E5C42D1D2FB12B2681D31B11E71DAF94400FB2E3774DB0CD44A4909C8CFB2D56E8A839FB3341991DF3FC89D8A73750FEF9A3BFC84A9C785AC7480A7548223C4AE471B6CD6FB01254A99383397F70454B4F177909CF330B030373EE669FEDE7C54677CE138803F3AE4571BAF597140B6FC5D61C9E55950A9E400AD03E530F693C869B8D89591EC71FEC0AB39F8076EFB60FEA71AD9&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=BA1E75DF0B7E0EB2&aid=5B63D875D0BD012B7BDA560C90DC37AE&vid=&iid=9CF7A0430CBB2DFD&sid=6D25DD85174CF6DB&eid=2A22E972FD97071B&fileno=20230903&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="BA1E75DF0B7E0EB2"; var my_aid="5B63D875D0BD012B7BDA560C90DC37AE";