Yang Xiaoyu,Zhao Yuqian,Yang Zhen,Li Shuzhou,Shao Qigang,Cao Ying.A dosimetric study on off-target isocenter plans for stereotactic body radiotherapy of lung cancer[J].Chinese Journal of Radiological Medicine and Protection,2023,43(8):633-638
A dosimetric study on off-target isocenter plans for stereotactic body radiotherapy of lung cancer
Received:February 14, 2023  
DOI:10.3760/cma.j.cn112271-20230214-00035
KeyWords:Lung cancer  SBRT  Off-target isocenter plan  Delivery accuracy
FundProject:国家自然科学基金(12005306,62076256);湖南省自然科学基金(2021JJ40960,2021JJ40966,2022JJ30976)
Author NameAffiliationE-mail
Yang Xiaoyu Oncology Department, National Clinical Research Center for Geriatric Disorders/Xiangya Hospital, Central South University, Changsha 410008, China  
Zhao Yuqian School of Automation, Central South University, Changsha 410083, China  
Yang Zhen Oncology Department, National Clinical Research Center for Geriatric Disorders/Xiangya Hospital, Central South University, Changsha 410008, China  
Li Shuzhou Oncology Department, National Clinical Research Center for Geriatric Disorders/Xiangya Hospital, Central South University, Changsha 410008, China  
Shao Qigang Oncology Department, National Clinical Research Center for Geriatric Disorders/Xiangya Hospital, Central South University, Changsha 410008, China  
Cao Ying Oncology Department, National Clinical Research Center for Geriatric Disorders/Xiangya Hospital, Central South University, Changsha 410008, China yingcao@csu.edu.cn 
Hits: 901
Download times: 290
Abstract::
      Objective To investigate the effects of off-target isocenter plans with different off-target distances on the plan quality and delivery accuracy of stereotactic body radiotherapy (SBRT) for lung cancer, aiming to provide a reference for the clinical plan design of SBRT for lung cancer.Methods For 10 lung cancer patients treated with SBRT, isocenter reference plans were designed by setting the plan isocenters at the mass centers of tumors and 60 off-target isocenter plans by setting the isocenters at distances of 1, 3, 5, 8, and 10 cm from the mass centers of tumors. The dosimetric differences between the off-target isocenter plans and the reference plans. Subsequently was analyzed, under different positional errors (0-5 mm). The gamma pass rates (GPRs) of these plans were measured using the Octavius 4D high-resolution dose verification system, and 240 verifications of these plans were completed. The robustness of the delivery accuracy of the reference plans and off-target isocenter plans were analyzed under different positional errors.Results The off-target isocenter plans yielded slightly worse dose gradient indices than the isocenter reference plans, but there was no statistically significant differences. With an increase in the off-target distance, the mean lung dose (MLD), V20 of normal lungs, as well as the Dmax of bronchi, showed slight upward trends. Compared with the isocenter reference plans, the MLD of the off-target isocenter plans increased by 0.8%, 0.8%, and 1.9% at off-target distances of 1, 3, and 10 cm, respectively, with statistically significant differences (z =-2.34 to -1.99, P < 0.05), and the V20 of the off-target isocenter plans increased by 2.0%, 2.5%, and 3.7% at off-target distances of 1, 5, and 10 cm, respectively, with statistically significant differences (z =-2.11 to -1.99, P < 0.05). In the case of a positional error of up to 5 mm, the GPRs of plans with off-target distances of 5 cm and above decreased by more than 1.0% on average and up to a maximum of 3.5%, showing statistically significant differences (z=2.13-2.75, P < 0.05).Conclusions Compared to the reference plans, the off-target isocenter plans showed slightly lower dosimetric quality and less robust delivery accuracy under different positional errors. Therefore, it is necessary to avoid the plans and treatment with too large off-target distances (≥ 5 cm) as far as possible for SBRT of lung cancer.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9061536  On-line:0

v
Scan QR Code
&et=0C253F3420628768D34E7B5B5512094159851FC0CDAAA2E40216DC0966C3F2E1AA1637DDE430B82566C1C9A9178B77C772740667B4CF1324B4D7CDF205108C52C529FDB66033B6B9F2342EC19181E0B1D17698ECCE4EC644042E58A32692D00B6C0257B51CE3EF1A&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=BA1E75DF0B7E0EB2&aid=55225852A69E5F8BB47175181AAA215B&vid=&iid=5D311CA918CA9A03&sid=06F643376BC2509E&eid=50B6AC44200581A5&fileno=20230809&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="BA1E75DF0B7E0EB2"; var my_aid="55225852A69E5F8BB47175181AAA215B";