Tian Shuang,Han Xiaole,Kong Xiangchuang,Lei Ziqiao,Yu Jianming,Li Cuiling.Clinical application of bedside X-ray photography assistor in neonates[J].Chinese Journal of Radiological Medicine and Protection,2023,43(4):297-300
Clinical application of bedside X-ray photography assistor in neonates
Received:November 22, 2022  
DOI:10.3760/cma.j.cn112271-20221122-00455
KeyWords:Bedside X-ray  Radiation dose  Neonate  Protection
FundProject:
Author NameAffiliationE-mail
Tian Shuang Department of Pain, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430022, China  
Han Xiaole Department of Radiology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Hubei Key Laboratory of Molecular Imaging, Wuhan 430022, China  
Kong Xiangchuang Department of Radiology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Hubei Key Laboratory of Molecular Imaging, Wuhan 430022, China  
Lei Ziqiao Department of Radiology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Hubei Key Laboratory of Molecular Imaging, Wuhan 430022, China  
Yu Jianming Department of Radiology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Hubei Key Laboratory of Molecular Imaging, Wuhan 430022, China  
Li Cuiling Wuhan Prevention and Treatment Center for Occupational Diseases, Wuhan 430015, China 8430051@qq.com 
Hits: 1197
Download times: 502
Abstract::
      Objective To assess the clinical application of bedside X-ray photography assistor (Patent No. 202023219898.1) in neonatal bedside photography.Methods From April 2021 to February 2022, a total of 180 pediatric patients were selected who underwent bedside chest X-ray photography in neonatal intensive care unit (NICU) of Union Hospital Affiliated to Tongji Medical College of Huazhong University of Science and Technology. These patients were divided into contrpol group, consisting of 48 males and 42 females aged at (3.3 ±2.0) d (0-10 d), and experimental group, including 50 males and 40 females aged (3.1±2.2) d (0-12 d). For chest photography, routine workflow was followed in the control group while in experimental group bedside photography protection and body position fixing device was used. The examination time, reshoot rate and image quality were compared between the two groups.Results The diagnostic physician score and patient comfort score in the experimental group were higher than those in the control group, with statistically significant differences (t = 3.98, 3.82, P < 0.001). The success rate in the experimental group was higher than that in the control group, with statistically significant difference (χ2= 7.84, P< 0.05). The average time of examination in the experimental group was not significantly different from in the control group (P>0.05 ).Conclusions The application of bedside X-ray photography assistor in neonatal bedside photography can significantly improve the success rate and image quality and reduce the radiation dose to pediatric patients without significantly increased examination time, which is worthy of clinical application and promotion.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9074807  On-line:0

v
Scan QR Code
&et=3C06DB566315F8DCA49CCF62F164EBF825D55634B5F9D73C6B3E64B0E7BC1A4AE5960A5E4E4B6A06830A54E4426C82B6B498980E5EE9A5FF98CE648B7CDDB4B7F3AADFAC5DBA8963&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=BA1E75DF0B7E0EB2&aid=7C6F8F037643673179A165A599675182&vid=&iid=E158A972A605785F&sid=E42CAFB11D4BE21A&eid=B1E36BF7B9783A85&fileno=20230409&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="BA1E75DF0B7E0EB2"; var my_aid="7C6F8F037643673179A165A599675182";