Zhou Qiang,Zhang Jing,Yang Baolu,Li Wenhong,Li Zeshu,Tuo Fei.Study on influencing factors associated with the analysis of 226Ra gamma spectra[J].Chinese Journal of Radiological Medicine and Protection,2023,43(2):112-117
Study on influencing factors associated with the analysis of 226Ra gamma spectra
Received:September 13, 2022  
DOI:10.3760/cma.j.cn112271-20220913-00372
KeyWords:γ-spectrometer  226Ra  222Rn
FundProject:
Author NameAffiliationE-mail
Zhou Qiang Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Zhang Jing Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Yang Baolu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
孔令南   
Li Wenhong Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Li Zeshu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Tuo Fei Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China flytuo@163.com 
Hits: 1086
Download times: 299
Abstract::
      Objective To measure the influencing factors associated with the analysis of 226Ra gamma spectra in order to obtain more acurate and precise measuring result on a basis of further optimizing the gamma spectrometry method for 226Ra activity.Methods A laboratory-based HPGe gamma spectrometer was used to carry out the studies on tracking measurement of sample sealing time, measurement of background fluctuation with lead shielding, analytical method, and selection of characteristic gamma ray energy peak of its daughter nuclides in 226Ra measurement.Results After the sample was sealed for 12 d, the decay products of 226Ra-222Rn basically reached equilibrium. The day and night fluctuations of 222Rn in the shielded lead room were obvious but had no obvious regularity. The way of filling nitrogen into the shielded lead room could reduce or avoid the influence of background fluctuations. For 31 soil samples measured after 23 days of sealing, the result of using the efficiency curve method showed that the 226Ra result calculated from the 351.9 keV energy peak were generally higher than the 609.3 keV energy peak, and the higher ratio ranged from 8.0% to 20.7%. The result of relative comparison method showed that the deviation ratio of the two peaks ranged from -4.1% to 10.3%.Conclusions It is recommended to consider the uncertainty attributed from decay equilibrium about 4% of measured at 12 d after the sample is sealed. When filling nitrogen through the shielded lead chamber to avoid background fluctuations, attention must be paid to the matching relationship between the volume of the lead chamber and the nitrogen filling flow. When the efficiency curve method is used to analyze the activity of 226Ra, the 214Bi (609.3 keV) energy peak has the effect of cascade coincidence addition, so 214Bi(609.3 keV) energy peak should be avoided. When the relative comparison method is used to analyze the activity of 226Ra, both two energy peaks of 214Pb (351.9 keV) and 214Bi (609.3 keV) can be used.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9069264  On-line:0

v
Scan QR Code
&et=08EE5A5866ACE4406591FD2C217E1FE371655143D75E974595C99EC6E79E6B8E462587BA69E45E4B85ADB10DA0AA5B9BE530DF43431F55DAB3AB5BD82D40071F81D724547297EFD08A837CAC4F86901A4A61A214F822FCB3961EC7D858FA7442&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=BA1E75DF0B7E0EB2&aid=F7D0025B820CBB4BA6B5762634CFCDAD&vid=&iid=0B39A22176CE99FB&sid=4BB057F167CF3A60&eid=7555FB9CC973F695&fileno=20230206&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="BA1E75DF0B7E0EB2"; var my_aid="F7D0025B820CBB4BA6B5762634CFCDAD";