Mao Xu,Xu Hui,Cheng Jinsheng,et al.Analysis of test results of radiological protection in workplace of radiological diagnosis and treatment institutions in south Xinjiang from 2018 to 2021[J].Chinese Journal of Radiological Medicine and Protection,2022,42(6):454-459
Analysis of test results of radiological protection in workplace of radiological diagnosis and treatment institutions in south Xinjiang from 2018 to 2021
Received:April 12, 2022  
DOI:10.3760/cma.j.cn112271-20220412-00158
KeyWords:X-ray diagnostic equipment  Radiological protection  Equipment test
FundProject:
Author NameAffiliationE-mail
Mao Xu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Xu Hui Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China xuhui@nirp.chinacdc.cn 
Cheng Jinsheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Liu Guoqing Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Zhang Jiaxin Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Cao Jianbiao Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Wang Yuwen The Center for Disease Control and Prevention of XinJiang Uygur Autonomous Region, Urumqi 830002, China  
Zhou Xiaolan Inner Mongolia Autonomous Region Comprehensive Center for Disease Control and Prevention, Huhhot 010031, China  
Zhou Xiaolan Kashgar Region Center for Disease Control and Prevention, Kashgar 844099, China  
Han Feng The National Institute for Occupational Health and Poison Control, Chinese Center for Disease Control and Prevention, Beijing 100050, China  
Hits: 3385
Download times: 1541
Abstract::
      Objective To analyze the radiological protection situation in the workplace of medical X-ray diagnostic equipment in primary medical institutions in south Xinjiang Uygur Autonomous Region, and then put forward necessary measures and suggestions.Methods In accordance with the national medical radiation protection monitoring program and the requirements of relevant standards for radiological health, medical X-ray diagnostic equipment in radialogical diagnosis and treatment institutions was tested selectively for workplace radiological protection, with the result statistically analyzed.Results From 2018 to 2021, radiological protection test was conducted for 84 workplaces in 15 radiological diagnosis and treatment institutions in 4 districts of south Xinjiang, with a pass rate of 98.8%, which was consistent with the mainland including the eastern and central regions in the country.Conclusions Based on the current situation in radiological diagnosis and treatment institutions on medical radiation protection in south Xinjiang, it is recommended to strengthen the supervision and management of radiological diagnosis and treatment equipment, improve the testing and technical capabilities of the local medical radiation monitoring technical teams and raise the level of radiological health work in the south Xinjiang.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12029804  On-line:0

v
Scan QR Code
&et=6FD48CA0A0DD475950D41E030AE6A6D73BCB643612A620A0D67FE3D7FC113CF236A8DE65055E49D22A2BABF3072319D6480A953CD6B10DA3A43207A87D497FA1F538F5CECDEF8D8F58D204D08C2374046C3F4CEC8256F0C3248BC246D55C6A9C7622BF56BCEF931FB1F9354B4508938E48942B7BA5473DA9D49722F968D531BC3E7CCA4EDC6D89501FE1696A56875398B6CBCC08FF73F06092A8E87F65AFFD96&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=885CEFEC57DA488F&aid=19AD2818E2D99015402DEDF635CB85B5&vid=&iid=B31275AF3241DB2D&sid=B1989ED92BA7E896&eid=3389C664025A98F9&fileno=20220610&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="885CEFEC57DA488F"; var my_aid="19AD2818E2D99015402DEDF635CB85B5";