Guo Senlin,Ren Yue,Zhang Yongxian,Zhang Zongrui,Yu Binbin,Liu Dandan,Niu Yantao.The experimental study of the effect of detector combination on image resolution in wide-detector CT[J].Chinese Journal of Radiological Medicine and Protection,2022,42(5):396-399
The experimental study of the effect of detector combination on image resolution in wide-detector CT
Received:February 17, 2022  
DOI:10.3760/cma.j.cn112271-20220217-00058
KeyWords:Wide-detector CT  Contrast resolution  Scanning mode
FundProject:
Author NameAffiliationE-mail
Guo Senlin Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Ren Yue Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Zhang Yongxian Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Zhang Zongrui Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Yu Binbin Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Liu Dandan Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Niu Yantao Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China ytniu163@163.com 
Hits: 1720
Download times: 625
Abstract::
      Objective To investigate the effect of different scanning modes,detector width and location in detector on high and low contrast resolution of wide-detector CT image.Methods The Catphan600 phantom with high and low contrast resolution modules was scanned with GE Revolution CT at the same CTDIvol. The scans were performed with the detector widths of 40,80 and 160mm for sequential scanning mode and with the detector width/pitch combinations of 40 mm/0.516,40 mm/0.984,80 mm/0.508 and 80 mm/0.992 for spiral scanning mode. The resolution modules were placed at the adjacent region between two sequential scans,central and foot side edge in the longitudinal scanning range seperately. The subjective evaluation of the high and low contrast resolution was performed by two radiologists.Results The high contrast resolution was 8 LP/cm at adjacent region between two sequential scans with the detector width of 80mm or 160mm in sequential scanning mode, and at the pitch of 0.5 in spiral scanning mode, while it was 7 LP/cm for the rest of detector combinations. The distinguishable diameter was 3mm at 1% low contrast resolution at foot side edge with the detector widths of 80mm or 160mm in the sequential scanning mode, and it was 2mm for all the other conditions. The distinguishable diameter was 2mm at 1% low contrast resolution with the detector width of 40mm and pitch 0.516 in the spiral scanning mode and it was worse with the wider detector and larger pitch.Conclusions For the wide-detector CT, scanning mode,detector width,location in detector and pitches will affect the high and low contrast resolution to some degree. Appropriate selection should be done according to actual needs in clinical practice.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9056458  On-line:0

v
Scan QR Code
&et=3CDA020A755F89DBEAECC647567B5490991912CF511FA999F8792620DB6D5E5A5C2DCABB7F4B8981E9745A0BA6EBB9472F147CD419D98FA33D7728BFEE799EAEADF80E66543ED3A5AC9AFA38DA1EE04CF5993BA452174D06ACFC22D02BECF47F387481250E1623D7&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=885CEFEC57DA488F&aid=5400E811F3EF1387273CC4D31D771DB7&vid=&iid=94C357A881DFC066&sid=7D6CD8918B045FD4&eid=87545994EC2C1F12&fileno=20220511&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="885CEFEC57DA488F"; var my_aid="5400E811F3EF1387273CC4D31D771DB7";