Ma Xiufeng,Liu Qingyun,Gan Ruilin,Wang Zhe.Determination of gross α and gross β radioactivity in IAEA-2020-intercomparison samples[J].Chinese Journal of Radiological Medicine and Protection,2022,42(1):50-54
Determination of gross α and gross β radioactivity in IAEA-2020-intercomparison samples
Received:September 16, 2021  
DOI:10.3760/cma.j.cn112271-20210916-00380
KeyWords:Standard material  Thin source method  Thick source method  Evaporation
FundProject:
Author NameAffiliation
Ma Xiufeng Beijing Nuclear and Radiation Safety Center, Beijing 100089, China 
Liu Qingyun Beijing Nuclear and Radiation Safety Center, Beijing 100089, China 
Gan Ruilin Beijing Nuclear and Radiation Safety Center, Beijing 100089, China 
Wang Zhe Beijing Nuclear and Radiation Safety Center, Beijing 100089, China 
Hits: 1980
Download times: 902
Abstract::
      Objective To summarize the practices in measuring the gross α and gross β radioactivity in IAEA-2020-intercomparison samples, which could be expected to be beneficial to the similar analysis and research.Methods With 241Am, Th, 90Sr/90Y, 40K and 137Cs solution as standard materials, the gross α and gross β radioactivity in water and aerosol samples were determined using thin source method. With 241Am, 40K powder as standard materials, the gross α and gross β radioactivity in water and fish samples were measured using thick source method combined with evaporation.Results The result showed that the relative deviation and Z-score by using thin source method were 4.12%-31.6% and 0.14-1.71, respectively, and those from thick source combined with evaporation were 2.63%-32.5% and 0.11-0.93, respectively, with the acceptance rate being 100%.Conclusions Generally, standard material shall be selected in the same types of radionuclides and energies as in samples. The thin source method is appropriate for emergency monitoring in the event of an accident. The thick source combined with evaporation should be perfered to the environmental monitoring or the analysis of unknown samples in laboratory. And then an intercomparison should be done with thin source method based on the radioactivity in samples. This work can provide a technical reference for similar measurements.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9071470  On-line:0

v
Scan QR Code
&et=9CD5B09F01CCAFB9E145BD893230AA943E2CDB9051ED168DD4F414B6160C5D75723A2AF796652F6FE81FB748ED38759C76760117EC839D73FBB196EB90296300227E3B638961B386F22EAE6B36F75FB33A6513AF1651163EA0F3EE6581C709BB&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=885CEFEC57DA488F&aid=BAE5A74472E7046F5D7B3088B37CEE3D&vid=&iid=CA4FD0336C81A37A&sid=771152D1ADC1C0EB&eid=318E4CC20AED4940&fileno=20220109&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="885CEFEC57DA488F"; var my_aid="BAE5A74472E7046F5D7B3088B37CEE3D";