Liu Hui,Tan Zhan,Yao Jie,et al.Test and discussion on SPECT performance under the Chinese health industry standard WS523-2019[J].Chinese Journal of Radiological Medicine and Protection,2021,41(7):534-538
Test and discussion on SPECT performance under the Chinese health industry standard WS523-2019
Received:January 15, 2020  
DOI:10.3760/cma.j.issn.0254-5098.2021.07.011
KeyWords:Single photon emission computed tomography  Intrinsic uniformity  Intrinsic spatial linearity
FundProject:国家自然基金(81773364);四川省自然基金重点项目(2017SZ0004);全军后勤科研重大项目(AWS17J007)
Author NameAffiliation
Liu Hui Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, Notional Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Tan Zhan Guangdong Province Hospital for Occupational Disease Prevention and Treatment, Guangzhou 510300, China 
Yao Jie Shanghai Municipal Center for Disease Control & Prevention, 200336, China 
Zhai Hezheng Institute of Radiation Medicine, Chinese Academy of Medical Sciences, Tianjin 300192, China 
Chen Zhangfan Institute for Radiation Hygiene Protection, Guangxi Center for Disease Prevention and Control, Nanning 530028, China 
Zhang Ling Anhui Prevention and Treatment Center for Occupational Disease, Hefei 230041, China 
Wang Feng Department of Nuclear Medicine, Beijing Cancer Hospital, 100036, China 
Li Honglei Department of Nuclear Medicine, China-Japan Friendship Hospital, 100029, China 
He Hongtao Philips (China) Investment Co., Ltd, Beijing 100600, China 
Li Wangyuan GE Medical System, Shanghai Ltd., 201203, China 
Zhang Shuai Siemens Healthineers Digital Technology (Shanghai) Co., Ltd., 200126, China 
Geng Jianhua Department of Nuclear Medicine, Cancer Hospital, Chinese Academy of Medical Sciences, 100021, China 
Hits: 4280
Download times: 1706
Abstract::
      Objective To discuss the problems existing in the implementation of the Chinese health industry standard WS 523-2019 by testing SPECT device. Methods Under the WS 523-2019 standards, a total of 10 SPECT devices were tested with regard to their SPECT reconstructed spatial resolution (SRSR), system planar sensitivity (SPS), system spatial resolution (SSR), whole body system spatial resolution (WSSR), intrinsic uniformity (IU), intrinsic count rate performance (ICR), intrinsic spatial resolution (ISR) and intrinsic spatial linearity(ISL).Result Under the requirements of WS 523-2019 standards for qualified limits, there are 3 devices with ISL unqualified and the rest of the performances qualified. The new standards basically can meet the clinical requirements and reflected the overall performance of SPECT. Conclusions The distance between the radiation source and the surface of the detector has great influence on the spatial resolution.In the measurement of ISL, there must be a lead grid separately in the x and y directions. The lead grids with the parallel slits shall be positioned on the detector with the center slit centered on the detector. It is suggested to add rotation center in the new standards.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12031822  On-line:0

v
Scan QR Code
&et=17F9701B371991039E473797EA510E997D9F7179CFE8E39688607FE83306B62D3F6B8B0D44574D839CAB9A27D4AF785971745317021CEF72C7D830F33DE997A33BBB4085F51722FC69FA4CFE5439C491EBEA800D92B4D4B7B6573C0151275CA0&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9475FABC7A03F4AB&aid=F7D79C0C8A4ABF6B9183EFA89DC34DD1&vid=&iid=DF92D298D3FF1E6E&sid=D45762219109E903&eid=E49BED2EA9A8956B&fileno=20210711&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9475FABC7A03F4AB"; var my_aid="F7D79C0C8A4ABF6B9183EFA89DC34DD1";