Zhou Qiang,Han Yanqing,Zhang Jing,et al.A review of intercomparison of measurement results by gamma-ray spectrometry in three laboratories in China and Japan[J].Chinese Journal of Radiological Medicine and Protection,2020,40(10):758-764
A review of intercomparison of measurement results by gamma-ray spectrometry in three laboratories in China and Japan
Received:April 08, 2020  
DOI:10.3760/cma.j.issn.0254-5098.2020.10.005
KeyWords:γ-spectrometry  Radionuclide  Intercomparison
FundProject:国家自然科学基金(21806156)
Author NameAffiliationE-mail
Zhou Qiang Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Han Yanqing Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Zhang Jing Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Yang Baolu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Li Wenhong Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Li Zeshu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Tuo Fei Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China flytuo@163.com 
Hits: 2615
Download times: 1347
Abstract::
      Objective To improve the radionuclide measurement technology by γ-spectrometry throuth a review of the intercomparisons between China and Japan results over the past 20 years. Methods Before the beginning of each intercomparison, one of participating laboratories was agreed to be responsible for preparing the samples for the intercomparison. By intercomparison of the measurement results arnong the laboratories some potential problems during the sample pre-trement or analysis were found and solved. Results The intercomparisons had been carried out for 11 times over the past 20 years. Totally 290 measured and analyzed results had been obtained from 20 samples of six categories. The proportion 95.5% of measurement results were in agreement with each other within the deviation range. Conclusions The measuring and analytical capability of the laboratories involved were tested. The intercomparisons also provide long-term quality assurance measures for the radionuclide measurement and analysis in each laboratory.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12031581  On-line:0

v
Scan QR Code
&et=C080D1B1BFBE6EFC1021951D73C89E89B6F61053D54E0A4065695CB80527E302E507569149FCEBBF74CE507CB16E22C30639709F87FA5C637A4F5F48777053C831B45D241D9005357A084FD4AD18187D45346D053A7AA89F74A210DC3863E7A953674DD9CA751040493E725A1DC01CE0403D6BC594772015&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=0D1D160AB8016934&aid=5BDB0152C50CB1983B46A8186A170A69&vid=&iid=F3090AE9B60B7ED1&sid=3D8AB54CA690066A&eid=231F9A307C169827&fileno=20201005&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="0D1D160AB8016934"; var my_aid="5BDB0152C50CB1983B46A8186A170A69";