Feasibility study of using two dimensional array ion-chamber to verifiy relative dose distribution calculated with Acuros BV |
Received:May 16, 2019 |
DOI:10.3760/cma.j.issn.0254-5098.2020.01.011 |
KeyWords:2D array ion-chamber Acuros BV Brachy therapy |
FundProject: |
Author Name | Affiliation | E-mail | Zhao Qiang | Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China | | Wu Xiangyang | Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China | 13087566100@163.com | Chang Xiaobin | Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China | | Feng Tao | Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China | | Yang Di | Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China | | Zhang Kun | Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China | | Qu Ximei | Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China | | Wang Xuemin | Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China | | Deng Jia | Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China | |
|
Hits: 2095 |
Download times: 1235 |
Abstract:: |
Objective To study the feasibility of using the PTW729 2D array ion-chamber to verify the relative dose distribution calculated with the Acuros BV algorithm. Both advantages and disadvantages of the method were analyzed to provide reference for practical clinical practices. Methods Based on self-built measurement phantoms, the dose distribution on the same slice of the phantom was measured with PTW729 and film, respectively, under the same measurement condition and plan. The dose distributions obtained by the two method were compared with the result calculated with Acuros BV, separately, by using γ analytical tool. And the stability of the PTW729 was tested. Results The γ comparison value was 95.9% between the film and Acuros BV, 98.9% between the PTW729 and Acuros BV and 88.0% between the film and PTW729, with 95.0%, 100.0%, and 100.0%, in their stability test respectively. Conclusions PTW729 2D array ion-chamber can be applied to the rapid verification of Acuros BV algorithm-calculated relative dose distribution. |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|