Zhao Qiang,Wu Xiangyang,Chang Xiaobin,Feng Tao,Yang Di,Zhang Kun,Qu Ximei,Wang Xuemin,Deng Jia.Feasibility study of using two dimensional array ion-chamber to verifiy relative dose distribution calculated with Acuros BV[J].Chinese Journal of Radiological Medicine and Protection,2020,40(1):59-63
Feasibility study of using two dimensional array ion-chamber to verifiy relative dose distribution calculated with Acuros BV
Received:May 16, 2019  
DOI:10.3760/cma.j.issn.0254-5098.2020.01.011
KeyWords:2D array ion-chamber  Acuros BV  Brachy therapy
FundProject:
Author NameAffiliationE-mail
Zhao Qiang Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China  
Wu Xiangyang Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China 13087566100@163.com 
Chang Xiaobin Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China  
Feng Tao Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China  
Yang Di Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China  
Zhang Kun Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China  
Qu Ximei Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China  
Wang Xuemin Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China  
Deng Jia Department of Radiotherapy, Shaanxi Provincial Cancer Hospital, Xi'an 710061, China  
Hits: 2095
Download times: 1235
Abstract::
      Objective To study the feasibility of using the PTW729 2D array ion-chamber to verify the relative dose distribution calculated with the Acuros BV algorithm. Both advantages and disadvantages of the method were analyzed to provide reference for practical clinical practices. Methods Based on self-built measurement phantoms, the dose distribution on the same slice of the phantom was measured with PTW729 and film, respectively, under the same measurement condition and plan. The dose distributions obtained by the two method were compared with the result calculated with Acuros BV, separately, by using γ analytical tool. And the stability of the PTW729 was tested. Results The γ comparison value was 95.9% between the film and Acuros BV, 98.9% between the PTW729 and Acuros BV and 88.0% between the film and PTW729, with 95.0%, 100.0%, and 100.0%, in their stability test respectively. Conclusions PTW729 2D array ion-chamber can be applied to the rapid verification of Acuros BV algorithm-calculated relative dose distribution.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9057615  On-line:0

v
Scan QR Code
&et=08BA79BB04C4219F9D0A59D56E3175C4A7FDD7A22B89E5AC1F32D37B00318B77D570233EE385388618149F2D53FAF2629BC3315917F03B960BFFD2C34419FDB53AC73ADCFD17202D884CE63CE54760CAD0532FF205D1592D6B867F74E3BB6F88AC0FB47CE2D133365E61D6045D024F694E076CC0F1BF08801DAE35E102BAE845&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=0D1D160AB8016934&aid=A7840A837AAEE54EB29BA7AB99645511&vid=&iid=CA4FD0336C81A37A&sid=6AC2A205FBB0EF23&eid=E84BBBDDD74F497C&fileno=20200111&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="0D1D160AB8016934"; var my_aid="A7840A837AAEE54EB29BA7AB99645511";