Hu Qiaoqiao,Jiang Fan,Wang Meijiao,Zhang Yibao,Yao Kaining,Gong Jian,Wu Hao,Yue Haizhen.A comparative study of dosimetric verification of HalcyonTM treatment plans using Varian Portal Dosimetry and PTW array 1500[J].Chinese Journal of Radiological Medicine and Protection,2019,39(1):58-62 |
A comparative study of dosimetric verification of HalcyonTM treatment plans using Varian Portal Dosimetry and PTW array 1500 |
Received:June 06, 2018 |
DOI:10.3760/cma.j.issn.0254-5098.2019.01.011 |
KeyWords:HalcyonTM accelerator Dose γ 2D analysis Electronic portal imaging device |
FundProject:北京市自然科学基金(1174016、7172048、1184014);首都卫生发展科研专项(首发2018-4-1027);四川省科技厅项目(2018HH0099) |
Author Name | Affiliation | E-mail | Hu Qiaoqiao | Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education), Department of Radiotherapy, Peking University Cancer Hospital & Institute, Beijing 100142, China | | Jiang Fan | Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education), Department of Radiotherapy, Peking University Cancer Hospital & Institute, Beijing 100142, China | | Wang Meijiao | Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education), Department of Radiotherapy, Peking University Cancer Hospital & Institute, Beijing 100142, China | | Zhang Yibao | Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education), Department of Radiotherapy, Peking University Cancer Hospital & Institute, Beijing 100142, China | | Yao Kaining | Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education), Department of Radiotherapy, Peking University Cancer Hospital & Institute, Beijing 100142, China | | Gong Jian | Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education), Department of Radiotherapy, Peking University Cancer Hospital & Institute, Beijing 100142, China | | Wu Hao | Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education), Department of Radiotherapy, Peking University Cancer Hospital & Institute, Beijing 100142, China | | Yue Haizhen | Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education), Department of Radiotherapy, Peking University Cancer Hospital & Institute, Beijing 100142, China | yuehzh@163.com |
|
Hits: 2913 |
Download times: 1593 |
Abstract:: |
Objective To compare two pretreatment plan QA methods for HalcyonTM accelerator using Portal Dosimetry (PD), and PTW OCTAVIUS 1500 detector array + Octagonal phantom (Oct 1500) respectively.Methods Parallel measurement-based pretreatment QA using two methods was performed for 22 IMRT/VMAT plans (74 fields) that have been used to treat 20 patients recruited in the Halcyon clinical trial. Several γ 2D comparisons were also applied to provide guidelines for Halcyon planning QA.Results Using Oct1500 method, the γ 2D passing rates for 74 fields in 22 Plans were 95.26±3.59, 95.01±3.62 (Local Dose), 99.05±1.35, 98.57±1.96 (Max Dose) respectively. Two-related samples non-parametric tests suggested that the differences between the evaluation criteria were of statistical significance (Z=-7.220, -4.108, P<0.05). For PD method, the γ 2D passing rates were 84.11%±1.35% (1 mm/1%), 99.07%±1.35% (2 mm/2%), and 99.86%±1.35% (3 mm/3%). Two-related samples non-parametric tests suggested that the differences between evaluation criteria of PD method were statistically significant (Z=-7.475, -7.475, -6.906, P<0.05). For 74 fields and max dose, 3 mm/3% evaluation criteria, two-related samples non-parametric tests suggested that the differences between PD and Oct1500 method were statistically significant (Z=-5.072, P<0.05).Conclusions Both methods can be used for Halcyon pretreatment plan QA. PD is superior to Oct1500 with respect to efficiency and spatial resolution-induced verification accuracy. The criteria of 2 mm/2% for PD, and Max Dose/3 mm/3% for Oct1500 was suggested. |