Qing Yunhua,Zhang Yan,Liu Jia,Zhao Shimin,Huang Lihua.Analysis of radioactivity in soil samples using γ-spectrometry in 2015 intercomparison[J].Chinese Journal of Radiological Medicine and Protection,2018,38(6):448-450 |
Analysis of radioactivity in soil samples using γ-spectrometry in 2015 intercomparison |
Received:November 05, 2017 |
DOI:10.3760/cma.j.issn.0254-5098.2018.06.010 |
KeyWords:HPGe γ-spectrometry Soil Radioactivity Intercomparison |
FundProject: |
Author Name | Affiliation | E-mail | Qing Yunhua | Fujian Center for Prevention and Control of Occupational Diseases and Chemical Poisoning, Fuzhou 350000, China | | Zhang Yan | Fujian Center for Prevention and Control of Occupational Diseases and Chemical Poisoning, Fuzhou 350000, China | | Liu Jia | Fujian Center for Prevention and Control of Occupational Diseases and Chemical Poisoning, Fuzhou 350000, China | | Zhao Shimin | Fujian Center for Prevention and Control of Occupational Diseases and Chemical Poisoning, Fuzhou 350000, China | | Huang Lihua | Fujian Center for Prevention and Control of Occupational Diseases and Chemical Poisoning, Fuzhou 350000, China | 2312933@qq.com |
|
Hits: 2366 |
Download times: 1413 |
Abstract:: |
Objective To analyze the specific activity of radionuclides based on laboratory intercomparison. Methods Intercomparison of the measurements of radioanuclides 232Th, 40K and 137Cs in soil samples, organized by National Institute for Radiological Protection, China CDC, was conducted using HPGe γ-spectrometry and the method of relative comparison and efficiency curve, respectively. Results The relative deviations were all less than 5% compared to the reference values in the conditions of both relative comparison method and efficiency curve method used, respectively. The relative deviations to the reference values were less than 4% for the relative comparison method. Conclusions Both the relative comparison method and the efficiency curve method can be used to measure the activity of radionuclides in the intercomparison samples, and the latter is superior to the former. |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|