Li Xuemin,Peng Ran,Jiang Yuliang,Ji Zhe,Guo Fuxin,Sun Haitao,Fan Jinghong,Li Xu,Li Weiyan,Wang Junjie.Dosimetric analysis of 3D-printing template assisted and CT-guided 125I seed implantation for treatment of soft tissue sarcoma[J].Chinese Journal of Radiological Medicine and Protection,2018,38(5):350-354
Dosimetric analysis of 3D-printing template assisted and CT-guided 125I seed implantation for treatment of soft tissue sarcoma
Received:October 31, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2018.05.005
KeyWords:Three-dimensional printing template  125I seeds  CT-guidance  Dosimetry  Soft tissue sarcoma
FundProject:
Author NameAffiliationE-mail
Li Xuemin Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Peng Ran Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Jiang Yuliang Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Ji Zhe Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Guo Fuxin Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Sun Haitao Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Fan Jinghong Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Li Xu Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Li Weiyan Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Wang Junjie Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China junjiewang_edu@sina.cn 
Hits: 2899
Download times: 1485
Abstract::
      Objective To compare the dose distribution of postoperative plans with preoperative plans of 3D-printing template (coplanar and non-coplanar) assisted and CT-guided 125I seed implantation for the treatment of soft tissue sarcoma, and to explore the accuracy of treatment at dosimetry level. Methods From December 2015 to July 2017, 19 patients with soft tissue sarcoma (a total of 25 lesions) were treated with 3D printing template assisted and CT-guided 125I seed implantation in Peking University Third Hospital.All patients underwent preoperative assessment, CT simulation orientation, preoperative planning, 3D-template printing, 3D-template reduction, needle and seed implantation, postoperative dosimetry assessment, postoperative care and follow-up. The preoperative and postoperative dosimetric parameters were compared. Ten cases of soft tissue sarcoma in superficial trunk or limbs were screened. Preoperative planning of coplanar template and non-coplanar template were designed respectively. The dosimetric parameters of preoperative planning guided of two templates were compared. Results Twenty-five 3D-printing templates were designed and constructed, and 25 lesions were totalled. There was no statistical difference between preoperative and postoperative dosimetric parameters. There was no statistical difference of the preoperative plan's dosimetric parameters between coplanar and non-coplanar in soft tissue sarcoma of superficial trunk/limbs. Conclusions The validation of actual dose distribution in postoperative plans assisted by 3D-printing template in 125I seed implantation showed that most of parameters could meet the expectation of preoperative plans, which indicated the improvement in accuracy for this new modality. For soft tissue sarcoma located in the superficial trunk/limbs, it was recommended to select the 3D-printing coplanar template firstly.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9060630  On-line:0

v
Scan QR Code
&et=8400690B225D0A443094374DB02750A4A96D787F0920509B126C3C22DB6211078460DB2B1072A9A0E14465CABAEBC4BADE91F6D5C62F1DD579595DF71EB2278EF814491B9894D34C05FE6F1B7D00F47DB029041B2C2EEAE1ABC03744A3F107CE6A5F3F2EDA6D32A398EB81CC24E0FA099A619422B6AD055F573489046DFA7E0E1F09DA1F5AE46E416DBF74A4141A1AFC&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=415CD46E06B7C80CE6C72F49054F1802&vid=&iid=94C357A881DFC066&sid=406BF8ED3BCE1927&eid=0954045FA0C6885F&fileno=20180505&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="415CD46E06B7C80CE6C72F49054F1802";