Zhang Fan,Wu Haibiao,Xiao Ainong,Ai Xiaohong,Cai Manbo,Li Pian,Cheng Lixiang,Zhang Zhenchao,Qu Guopu.Monte Carlo-based simulation of influence of linear accelerator beam parameter on percentage depth dose[J].Chinese Journal of Radiological Medicine and Protection,2018,38(2):145-149
Monte Carlo-based simulation of influence of linear accelerator beam parameter on percentage depth dose
Received:August 31, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2018.02.013
KeyWords:Monte Carlo  Electron energy  Radial intensity distribution  Percentage depth dose
FundProject:湖南省自然科学基金(2016JJ3108);湖南省重点研发计划项目(2015SK2022)
Author NameAffiliationE-mail
Zhang Fan School of Nuclear Science and Technology, University of South China, Hengyang 421001, China  
Wu Haibiao Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China  
Xiao Ainong Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China  
Ai Xiaohong Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China  
Cai Manbo Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China  
Li Pian Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China  
Cheng Lixiang Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China  
Zhang Zhenchao School of Nuclear Science and Technology, University of South China, Hengyang 421001, China  
Qu Guopu School of Nuclear Science and Technology, University of South China, Hengyang 421001, China quguopu@usc.edu.cn 
Hits: 2896
Download times: 1816
Abstract::
      Objective To determine the optimal electron beam energy at different field size through a Monte Carlo-based simulation of the therapy head of Varian IX 6 MV linear accelerator so as to study the influence of radial intensity on depth dose. Methods Firstly, keeping the radial intensity unchanged for the field of interest while changing electron beam energy, compassion was carried out of calculated percentage depth doses between measured values. Thus, the optimal energy was identified for this field size. Then, the obtained energy was set the optimal value to study the radial intensity influence on the depth doses. Results The optimal electron energy for 4 cm×4 cm, 10 cm×10 cm, 20 cm×20 cm and 30 cm×30 cm field sizes was 5.9, 6.0, 6.3 and 6.4 MeV respectively. Changes in radial intensities resulted in negligible changes in percentage depth doses for 4 cm×4 cm and 10 cm×10 cm fields, but led to observable discrepancy for 20 cm×20 cm and 30 cm×30 cm fields. Conclusions The optimal electron energies for different field sizes are slightly different. Change in radial intensity distribution has significant influence on the depth dose for large field. To improve simulation accuracy, the field size needs to be taken into consideration in determining the electron beam energy and radial intensity distribution.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9078197  On-line:0

v
Scan QR Code
&et=688C18234B3B6C89815CD904567836213B533EB8B427B73F5E4F433942CA6B186201820412D5DA296F0D7980EBE77C2781755902D48BF5B663EA773A4152A0A92E34D04AC49B62BCAF88E25631BD75367CD681C4C66130F9A629FE32025108DCC12366A95A314655&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=295B27E71BBC5E120939938D53F01762&vid=&iid=0B39A22176CE99FB&sid=769BD58726D66E7D&eid=EB552E4CFC85690B&fileno=20180213&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="295B27E71BBC5E120939938D53F01762";