Zhang Fan,Wu Haibiao,Xiao Ainong,Ai Xiaohong,Cai Manbo,Li Pian,Cheng Lixiang,Zhang Zhenchao,Qu Guopu.Monte Carlo-based simulation of influence of linear accelerator beam parameter on percentage depth dose[J].Chinese Journal of Radiological Medicine and Protection,2018,38(2):145-149 |
Monte Carlo-based simulation of influence of linear accelerator beam parameter on percentage depth dose |
Received:August 31, 2017 |
DOI:10.3760/cma.j.issn.0254-5098.2018.02.013 |
KeyWords:Monte Carlo Electron energy Radial intensity distribution Percentage depth dose |
FundProject:湖南省自然科学基金(2016JJ3108);湖南省重点研发计划项目(2015SK2022) |
Author Name | Affiliation | E-mail | Zhang Fan | School of Nuclear Science and Technology, University of South China, Hengyang 421001, China | | Wu Haibiao | Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China | | Xiao Ainong | Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China | | Ai Xiaohong | Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China | | Cai Manbo | Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China | | Li Pian | Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China | | Cheng Lixiang | Department of Oncology, First Hospital Affiliated to University of South China, Hengyang 421001, China | | Zhang Zhenchao | School of Nuclear Science and Technology, University of South China, Hengyang 421001, China | | Qu Guopu | School of Nuclear Science and Technology, University of South China, Hengyang 421001, China | quguopu@usc.edu.cn |
|
Hits: 2896 |
Download times: 1816 |
Abstract:: |
Objective To determine the optimal electron beam energy at different field size through a Monte Carlo-based simulation of the therapy head of Varian IX 6 MV linear accelerator so as to study the influence of radial intensity on depth dose. Methods Firstly, keeping the radial intensity unchanged for the field of interest while changing electron beam energy, compassion was carried out of calculated percentage depth doses between measured values. Thus, the optimal energy was identified for this field size. Then, the obtained energy was set the optimal value to study the radial intensity influence on the depth doses. Results The optimal electron energy for 4 cm×4 cm, 10 cm×10 cm, 20 cm×20 cm and 30 cm×30 cm field sizes was 5.9, 6.0, 6.3 and 6.4 MeV respectively. Changes in radial intensities resulted in negligible changes in percentage depth doses for 4 cm×4 cm and 10 cm×10 cm fields, but led to observable discrepancy for 20 cm×20 cm and 30 cm×30 cm fields. Conclusions The optimal electron energies for different field sizes are slightly different. Change in radial intensity distribution has significant influence on the depth dose for large field. To improve simulation accuracy, the field size needs to be taken into consideration in determining the electron beam energy and radial intensity distribution. |
HTML View Full Text View/Add Comment Download reader |