Sheng Yinxiangzi,Wang Weiwei,Huang Zhijie,Zhao Jingfang,Chien Hsi,Kambiz Shahnazi.Measurement of CT Hounsfield Units and relative stopping powers conversion curve in proton and carbon ion therapy[J].Chinese Journal of Radiological Medicine and Protection,2017,37(9):667-670
Measurement of CT Hounsfield Units and relative stopping powers conversion curve in proton and carbon ion therapy
Received:February 27, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2017.09.005
KeyWords:CT value  Relative stopping power  Ion therapy
FundProject:上海市青年科技英才扬帆计划(16YF1410700)。
Author NameAffiliationE-mail
Sheng Yinxiangzi Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Shanghai 201321, China  
Wang Weiwei Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Shanghai 201321, China  
Huang Zhijie Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Shanghai 201321, China  
Zhao Jingfang Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Shanghai 201321, China  
Chien Hsi Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Shanghai 201321, China  
Kambiz Shahnazi Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Shanghai 201321, China kambiz88224@yahoo.com 
Hits: 3516
Download times: 2116
Abstract::
      Objective To measure the CT Hounsfield Unit (HU) and relative stopping power (RSP) conversion curve.Methods In this study, the RSPs of 12 different tissue equivalent rods were measured with proton and carbon beam in the Shanghai Proton and Heavy Ion Center (SPHIC). The same tissue equivalent materials were scanned with CT scanner to acquire the HU.Results Conversion curve for the transformation of HU into RSP was generated for both proton and carbon ion beam. Differences between RSPs measured using proton and carbon beam were ≤0.64% except lung material.Conclusions A RSP versus HU conversion curve was generated for both protons and carbon ions.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9076679  On-line:0

v
Scan QR Code
&et=DC44477595533BA78A6A4E546A9263A59C4F245098850730B62F8FD0FB94FFEECCE81BF04F0A3F78AEC153874E9D14B54B8ED9B56B2CEA2683D64DC5E2259697AFE6E9E9A192A25CD434CEE2FD66E79B58778616248B06B40AFBBD4297ED863C49D3A79797D1EEB2F99DB87EF1D6FF14D7AB7487FFD700EC&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FA004A8A4ED1540B&aid=CF1332E5A7C16DE7E1A2B96F4FA7081A&vid=&iid=9CF7A0430CBB2DFD&sid=06DAE5E1DF7D0B6A&eid=2B6C525BCE31A7DA&fileno=20170906&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FA004A8A4ED1540B"; var my_aid="CF1332E5A7C16DE7E1A2B96F4FA7081A";