Ding Hailing,Wang Yongchun,Jiang Wei.Practicable value of DAP (dose-area product) database in occupational radiation protection in cardiovascular interventional procedures[J].Chinese Journal of Radiological Medicine and Protection,2017,37(8):623-625,629
Practicable value of DAP (dose-area product) database in occupational radiation protection in cardiovascular interventional procedures
Received:January 10, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2017.08.013
KeyWords:Dose area product  Cardiovascular intervention  Database
FundProject:
Author NameAffiliationE-mail
Ding Hailing Department of Radiology, Affiliated Changhai Hospital, Secondary Military Medical University, Shanghai 200433, China  
Wang Yongchun Department of Radiology, Affiliated Changhai Hospital, Secondary Military Medical University, Shanghai 200433, China  
Jiang Wei Department of Radiology, Affiliated Changhai Hospital, Secondary Military Medical University, Shanghai 200433, China 13916579426@126.com 
Hits: 2534
Download times: 1600
Abstract::
      Objective To evaluate the efficacy of the DAP database in reducing radiation hazards during cardiovascular intervention.Methods The study conducted a retrospective analysis of all consecutive cardiovascular cases in the database from April 1st,2016 to June 30th,2016.Based on the imaging systems,an image intensifier (Ⅱ) and a flat panel detector (FPD),the cases of patients undergoing coronary angiography (CAG) and percutaneous coronary interventions (PCI) were collected to observe Body Mass Index (BMI),fluoroscopy time (FT) and dose area product (DAP). Results The FT and DAP values in CAG were (3.26±2.59) min and (1 938±1 296)μGy·m2 for the Ⅱ system and (3.00±2.89) min and (3 718±2 859)μGy·m2 for the FPD system,respectively,whereas,the FT and DAP values in PCI were (17.81±12.71) min and (8 899±7 032)μGy·m2 for the Ⅱ system and (21.99±15.91) min and (19 526±14 134)μGy·m2 for the FPD system,respectively.The differences in DAP values (x±s) for CAG and PCI between the angiography systems were significant (t=-10.664,11.239,P<0.05).The DAP and total DAP values during PCI for fluoroscopy from the FPD system were 3 596 755.60 and 4 881 484.50 μGy·m2,respectively. Conclusions The DAP values in CAG and PCI using FPD system are much higher than the relevant data.The use of an FPD system resulted in higher DAP values compared to the Ⅱ system in cardiovascular interventional procedures.Reduction in fluoroscopy time may decrease radiation hazards during PCI.The DAP database allows for a fast,convenient and enough objective data to discovery timely difference and variation in dose in cardiovascular interventional procedures.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9056829  On-line:0

v
Scan QR Code
&et=0B21557D3534438A3FA11C45ACC06780A66F63FE2BDAE32EF8E867C03E7A326B979D15077C7318C4CF2DD74F70F83B58697809B5C0A18B80E829019CB7F25AA8F100B575671A00EA736C118F5B58B8FAE2D6356B5007D2CC605C63332CB67F2C79CF69CA7D7786C337A0C7CE4305AE03FD6C8E5E0ED627600F02572389BAB0CD4C383F0C11766DB4&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FA004A8A4ED1540B&aid=3C3EF3B75913B4FFC2F529257817878B&vid=&iid=5D311CA918CA9A03&sid=F8AEC975DBDD7F2F&eid=E46F4862E0752FB4&fileno=20170813&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FA004A8A4ED1540B"; var my_aid="3C3EF3B75913B4FFC2F529257817878B";