Jiang Yuliang,Ji Zhe,Guo Fuxin,Peng Ran,Sun Haitao,Fan Jinghong,Li Weiyan,Liu Kai,Lei Jinhua,Wang Junjie.Side effect of radioactive 125Ⅰ seed implantation for recurrent malignant tumor of head and neck assisted by 3D-printing individual guide plate[J].Chinese Journal of Radiological Medicine and Protection,2017,37(7):495-499
Side effect of radioactive 125Ⅰ seed implantation for recurrent malignant tumor of head and neck assisted by 3D-printing individual guide plate
Received:April 11, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2017.07.004
KeyWords:3D-printing non-coplanar template  Radioactive seed implantation  Side effect  Head and neck carcinoma
FundProject:首都临床特色应用研究与成果推广项目(Z151100004015171)
Author NameAffiliationE-mail
Jiang Yuliang Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Ji Zhe Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Guo Fuxin Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Peng Ran Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Sun Haitao Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Fan Jinghong Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Li Weiyan Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China  
Liu Kai Department of Oncology, Dingzhou City People's Hospital, Dingzhou 073000, China  
Lei Jinhua Department of Oncology, Shiyan City People's Hospital, Shiyan 442000, China  
Wang Junjie Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China junjiewang_edu@sina.cn 
Hits: 2475
Download times: 2029
Abstract::
      Objective To investigate the acute side effect of 3D printing individual non-coplanar template for 125Ⅰ seed implantation in head and neck recurrent/ metastatic carcinoma. Methods Between January 2016 and December 2016, 42 patients of local recurrent malignant tumor of head and neck received 3D printing guide plate assist radioactive seeds implantations, and included in the study. The tumor volume ranged from 2.4 to 102.8 cm3 (median 28.6 cm3). The prescribed dose is 110-160 Gy, and the seeds activity were 0.34 to 0.70 mCi(1 Ci=3.7×1010Bq). All patients carried out preoperative planning design, individual guide plate production, seed implantation, postoperative dose assessment, and follow-up. The side effects of skin, mucous membrane, blood and spinal cord were statistically analyzed. Results All patients were operated successfully. The follow-up time was 4-14 months (median 8.5 months). There were no adverse reactions at grade three or above. Three patients had grade one skin reaction. One patient experienced grade one mucosal reaction, two experienced grade two mucosal reactions. The skin response was correlated with the dose of the skin (χ2=7.067, P=0.032). No hematologic toxicity or radiation myelopathy was observed and no seed displacement was found. Conclusions 3D-printing guide plate can provide good accuracy for positioning and direction. For local recurrent malignant tumor of head and neck, there were no obvious adverse reactions and the operation was simple and the dosage was accurate.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9059984  On-line:0

v
Scan QR Code
&et=F21CBBD1BF6C4B3C1B8D9D133320B01F31B22178BE1A18C46FCBD8A2BFF0F75FC526577A9C339CECB65437086A354E6CECF7D3F684B26C8AC58FC54394CC94AD923D0FC032496396C1BFE54C33F11DDFCFE80637D638615ECC7826672F8FB213F9186FAE7DBE688BA95C266C87C50593A18FEC0483BC8A0FC256A59F423E4E55514B0A12551B477AE93EE17C803425C6B06D896D459F668057D834A21067FB2F&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FA004A8A4ED1540B&aid=BF88AFC87C5C19B434D67A3AF3408E32&vid=&iid=DF92D298D3FF1E6E&sid=DDDA4F26E8AD3C0E&eid=A03A15CF5604A8B0&fileno=20170704&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FA004A8A4ED1540B"; var my_aid="BF88AFC87C5C19B434D67A3AF3408E32";