Zhou Sheng,Chen Xiaofei,Cao Hongxia,Li Qinghua,Wang Xinzhu,Wang Wenqi.Experimental study of correlation between the focus-to-flat panel detector distance and entrance surface doses in digital radiography of lumbar spine[J].Chinese Journal of Radiological Medicine and Protection,2016,36(7):540-543
Experimental study of correlation between the focus-to-flat panel detector distance and entrance surface doses in digital radiography of lumbar spine
Received:December 20, 2015  
DOI:10.3760/cma.j.issn.0254-5098.2016.07.013
KeyWords:Radiography  Lumbar spine  Focus-to-flat panel detector distance  Entrance surface dose  Exposure index
FundProject:甘肃省卫生行业科研计划项目(GSWSKY-2014-12)
Author NameAffiliationE-mail
Zhou Sheng Department of Radiology, Gansu province Hospital of traditional Chinese medicin, Lanzhou 730050, China  
Chen Xiaofei Department of Radiology, Gansu province Hospital of traditional Chinese medicin, Lanzhou 730050, China cxf750915@sina.com 
Cao Hongxia Department of Radiology, Gansu province Hospital of traditional Chinese medicin, Lanzhou 730050, China  
Li Qinghua Department of Radiology, Gansu province Hospital of traditional Chinese medicin, Lanzhou 730050, China  
Wang Xinzhu Department of Radiology, Gansu province Hospital of traditional Chinese medicin, Lanzhou 730050, China  
Wang Wenqi Department of Radiology, Gansu province Hospital of traditional Chinese medicin, Lanzhou 730050, China  
Hits: 4083
Download times: 2194
Abstract::
      Objective To investigated the correlation between the focus-to-flat panel detector distance and entrance surface doses in digital radiography of lumbar spine. Methods The spine phantom used for bone mineral density(BMD)calibration was radiographed at different value of FFD (100~240 cm with the interval 10 cm,totally 15 times) using 70 kV and 90 kV with the automatic exposure control (AEC) mode without anti-scatter grid. The parameters measured by QUART dido 2100K type X-ray dose meter. So record the value parameters of the tube current (mAs) and entrance surface dose (ESD). The relation curve were counted between FFD, ESD, mAs and Exposure index (EI) use excle. Pearson correlation analysis and related coefficient difference analysis were performed using the SPSS 19.0 software (P < 0.05 for the statistically significant level). The FFD and ESD values of regression equation were obtained with regression analysis. Results (1) When the tube voltage was 70 kV: FFD and ESD were negative correlation (r=-0.922, P < 0.001), FFD and mAs were positive correlation (r=0.991, P < 0.001), FFD and EI negative correlation (r=-0.938, P < 0.001). (2) When the tube voltage was 90 kV: FFD and ESD were negative correlation (r=-0.955, P < 0.001), FFD and mAs were positive correlation (r=0.994, P < 0.001), FFD and EI showed no significant correlation (r=- 0.523, P > 0.05). (3) Different position correlation coefficient analysis of the difference was not statistically significant difference. (4) The regression analysis results: The regression equation of the lumbar anterior-posterior position FFD and the ESD at 70 kV: y=0.805 x + 319.731 (F=74.175, P <0.001).The regression equation of the lumbar lateral position FFD and the ESD at 90 kV:y=3.152 x + 457.613 (F=133.524, P <0.001).Conclusions The FFD and ESD in the lumbar spine X-ray radiography were relevant and highly negative correlation. It can effectively reduce the radiographic ESD of lumbar spine with the increasing of the FFD.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9069082  On-line:0

v
Scan QR Code
&et=3B5D9771435A916B26445566CEFC6D4B1CB73E7260244B150FF3E8BE9409F37BDCE122628D568343F71B1702F16B6B5C07D9665620C702D6FAD98E322EAB9BFC5DE4DDA14A2AAC1031B2D728C6CB38C0CC4A94AA546A8DB16A540DC9613167024DDB1196CBA22C62757BF0E58A514C1B8EF83D832B870CF4C66D103AC93019B8687C0450EEE0E9E0AD4FB1592F91546D73AA52F99F41FA3E&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9F915C6F01DE79C5&aid=65B039F9D05E9856D6405F401609D48A&vid=&iid=DF92D298D3FF1E6E&sid=DFBC046213B3DD86&eid=BA305A52E2EE9350&fileno=20160713&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9F915C6F01DE79C5"; var my_aid="65B039F9D05E9856D6405F401609D48A";