Liao Xiongfei,Li Churong,Li Jie,et al.Application of intensity-modulated radiotherapy for carniospinal irradiation with multi-ISO centers planning method[J].Chinese Journal of Radiological Medicine and Protection,2015,35(10):756-760
Application of intensity-modulated radiotherapy for carniospinal irradiation with multi-ISO centers planning method
Received:March 05, 2015  
DOI:10.3760/cma.j.issn.0254-5098.2015.10.008
KeyWords:Multi-ISO center planning method  Craniospinal system  IMRT  Helical tomotherapy
FundProject:
Author NameAffiliationE-mail
Liao Xiongfei Department of Radiation Oncology, Sichuan Cancer Hospital, Chengdu 610041, China  
Li Churong Department of Radiation Oncology, Sichuan Cancer Hospital, Chengdu 610041, China  
Li Jie Department of Radiation Oncology, Sichuan Cancer Hospital, Chengdu 610041, China  
Chen Yazheng Department of Radiation Oncology, Sichuan Cancer Hospital, Chengdu 610041, China  
Wang Pei Department of Radiation Oncology, Sichuan Cancer Hospital, Chengdu 610041, China dengwangpei@163.com 
Hits: 3587
Download times: 2796
Abstract::
      Objective To explore the application value of multi-ISO center planning method in intensity-modulated radiotherapy for carniospinal irradiation. Methods Ten patients treated with craniospinal irradiation with helical tomotherapy were selected. For these patients, new plans were designed with multi-ISO centers planning method on the treatment planning system (TPS) named Eclipse 11.0. Dose distribution to the tumor, OARs and normal tissue, the treatment time and the monitor units (MUs) of the two plans were compared. Results The plans designed in two groups satisfied all clinical requirements. For the tumor target (PTV), the difference of D95% between two groups was not statistical significant, while D98%, D2% and HI in HT group were superior to M-ISO group(t=2.822,2.333,4.743,P<0.05), D50%, Dmean and CI in M-ISO group were superior to HT group (t=5.259,3.685,8.835,P<0.05). The dose of OARs such as cochlea, parotid, submandibular gland, thyroid gland and kidney in HT group was lower than M-ISO group (t=4.365,5.416,2.674,3.077,2.782,2.607,4.659,P<0.05), and the dose of pancreas and small bowel was higher than M-ISO group(t=5.265,5.935,P<0.05). Differences were not significant for V5 of normal tissue between two groups; while V10, V20 and V36 of normal tissue in M-ISO group were lower than HT group (t=3.57,3.701,2.602,P<0.05). M-ISO group reduced 41.0% of the treatment time by average and reduced 94.1% MUs by average. Conclusions Intensity-modulated radiotherapy for carniospinal irradiation with multi-ISO centers planning method not only met the requirements of clinical dosimetry, but also shorten the treatment time, reduced the damage to the machine. Multi-ISO centers planning method might be promoted as a new design scheme.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12028918  On-line:0

v
Scan QR Code
&et=DCA5CABA83612A94527D8C68DFB10B6FB9A257F8E6988674B31735C0277AC5C60F2D5564949C5A174D011DF5AA09E1A86B1C7B6B3EAE2A06E917EDBDE9577264720FAC17305182FF56FC46305948BC0BCA7750378F3B420CFE1E4C7C9910E36A70A6E80068B3DFE1571ACAFCF97B5FF41B0CD2B5350569DA&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=6F3C2608457E700608006D62FED3CF7E&vid=&iid=F3090AE9B60B7ED1&sid=895BB10EB8F3BD49&eid=8D71AF42ACD39979&fileno=20151008&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="6F3C2608457E700608006D62FED3CF7E";