Yin Liangliang,Shao Xianzhang,Shen Baoming,Li Yucheng,Ji Yanqin.Determination of α-radionuclides in environmental samples by large area grid ionization chamber[J].Chinese Journal of Radiological Medicine and Protection,2015,35(6):461-464
Determination of α-radionuclides in environmental samples by large area grid ionization chamber
Received:January 30, 2015  
DOI:10.3760/cma.j.issn.0254-5098.2015.06.013
KeyWords:Grid ionization chamber  α-emitters  Environmental samples
FundProject:卫生行业科研专项资助项目(201002009);科技部科技支撑项目(2013BAK03B00)
Author NameAffiliationE-mail
Yin Liangliang Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Shao Xianzhang Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Shen Baoming Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Li Yucheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Ji Yanqin Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China jiyanqin@nirp.cn 
Hits: 2663
Download times: 2192
Abstract::
      Objective To develop the method of analyzing α-radionuclides using large area grid ionization chamber. Methods Ultrasonic dispersion and vacuum drying system was used to prepare sample source, large standard thin source and plutonium plane source were used to optimize the working condition of spectrometer, and calibrate the instrument for analyzing α emitters. The certified reference materials (GBW04127) were used to verify the accuracy of the method. Results The non-linearity of calibration curve for standard thin sources of neptunium, plutonium and americium was less than 0.2%, and the resolution were 112, 84 and 106 keV, respectively. The counting efficiency was 31.2% for the large standard thin source. The values of specific activity measured in this way were in good agreement with those of the certified reference materials. 232Th, 238U, 230Th, 234U/226Ra, 210Po, 222Rn and 218Po were analyzed in a uranium mineral sample, and their specific radioactivity values were 5.3, 3.8, 35.6, 21.4, 27.0, 19.6 and 11.1 Bq/g, respectively. Conclusions The method can be used to analyze α spectrum quickly in low-level radioactive environmental samples.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9075126  On-line:0

v
Scan QR Code
&et=9CD5B09F01CCAFB9E145BD893230AA948F56B75FCA73F0E9F0DA2EF4484B439B244AB3068623C39C4C0D63BB4C36B2BFFAB073A0E4F8832569E6B6868E2778D6153F693759C6490B15AFBF5425BE5CFDDD6E4F23D1C128B5CC4A71F30F7CACD6B06F206226ECF5C4&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=5F7737E5A04D36C7DBBBAC0892F13A06&vid=&iid=B31275AF3241DB2D&sid=FCACFF68346F8D4F&eid=1A033C02510EFBE6&fileno=20150613&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="5F7737E5A04D36C7DBBBAC0892F13A06";