Li Zhi,Jiang Shan,Yang Zhiyong,et al.Monte Carlo simulation and experimental investigation of 125I interseed dose attenuation[J].Chinese Journal of Radiological Medicine and Protection,2015,35(5):389-392 |
Monte Carlo simulation and experimental investigation of 125I interseed dose attenuation |
Received:September 03, 2014 |
DOI:10.3760/cma.j.issn.0254-5098.2015.05.019 |
KeyWords:Brachytherapy 125I seed dose attenuation Geant 4 Monte Carlo |
FundProject:国家自然科学基金(51175373);天津市科技支撑重大项目(12ZCDZSY10600) |
|
Hits: 3531 |
Download times: 2862 |
Abstract:: |
Objective To investigate the interseed dose attenuation for multiple 125I seeds. Methods Monte Carlo simulation was done by Geant 4 to investigate the dose distribution of single seed and multiple seeds. The results were compared with the dose calculation method from TG43-U1 and examined by experiments. Results The difference of single seed dose distribution between Monte Carlo method and line source model was ± 3%. The difference between Monte Carlo method and experiment result was ± 5%. The interseed dose attenuation of multiple seeds at the interesting points was 3.8% to 13.2% and the average interseed dose attenuation was 7.2%. The difference of experiment result and Monte Carlo result was less than 6%. Conclusions The interseed attenuation is about 7% and the maximum value may be larger than 13% for multiple seeds. The interseed dose attenuation may be larger in tissue. So it is not accurate to calculate the dose distribution by using TG43-U1. |