Zhang Jiyong,Wu Lili,Lu Jiayang,Zhang Wuzhe.Application of multi-leaf collimator quality assurance phantom on Truebeam[J].Chinese Journal of Radiological Medicine and Protection,2015,35(5):356-359
Application of multi-leaf collimator quality assurance phantom on Truebeam
Received:October 12, 2014  
DOI:10.3760/cma.j.issn.0254-5098.2015.05.009
KeyWords:Truebeam linear accelerator  Multi-leaf collimator  Quality assurance
FundProject:
Author NameAffiliationE-mail
Zhang Jiyong Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Shantou 515031, China  
Wu Lili Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Shantou 515031, China  
Lu Jiayang Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Shantou 515031, China  
Zhang Wuzhe Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Shantou 515031, China jonzh502@qq.com 
Hits: 2214
Download times: 2260
Abstract::
      Objective To perform a routine quality assurance procedure for Truebeam multi-leaf collimator (MLC) using MLC quality assurance (QA) phantom, verifying the reliability of MLC during the treatment. Methods MLC QA phantom was a specialized phantom for multi-leaf collimator QA, and contained five radio-opaque spheres that were embedded in an L-shape. The phantom was placed isocentrically on the Truebeam treatment couch for the tests. A quality assurance plan was settled up in the Eclipse v10.0 so that the fields for acquiring the necessary images could be created. The images were acquired by the electronic portal imaging device (EPID), and imported into the PIPSpro software for the analysis. The tests were delivered once a week for six weeks to verify consistency of the delivery, and the images were acquired in the same manner each time. Results For the leaf position test, the average position error was (0.21±0.02) mm. The leaf width was measured at the isocenter, and the average error was (0.04±0.02) mm for the leaf width test. Multi-Port test showed the dynamic leaf shift error, the average error was (0.26±0.04) mm. For the leaf transmission test, the average inter-leaf leakage value was 1.0%±0.14%. Conclusions The MLC system of Truebeam could operate very well and the QA phantom is a useful test tool for the MLC QA.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9074252  On-line:0

v
Scan QR Code
&et=DCA5CABA83612A94E99FEAC3692F26F42FA95909EC1026A1A4E280C4FAB7D5661C2765B7A364E2A38CC6C9C99C9D87078A84E35AA7FA5F24941C3963EDB56AFE766F8A62C2273ED89017C1CBF7371FD1&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=6C6113E1FCA18436A2C02EF8D87D954E&vid=&iid=94C357A881DFC066&sid=35E8A259891FB32F&eid=3356A7630A93A219&fileno=20150509&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="6C6113E1FCA18436A2C02EF8D87D954E";