Yang Jinlei,Liu Boyu,Lu Shuangchen,Bu Mingwei.Comparison of Monaco and Pinnacle treatment planning systems in volumetric modulated arc therapy optimization for lung cancer[J].Chinese Journal of Radiological Medicine and Protection,2015,35(2):119-122
Comparison of Monaco and Pinnacle treatment planning systems in volumetric modulated arc therapy optimization for lung cancer
Received:April 01, 2014  
DOI:10.3760/cma.j.issn.0254-5098.2015.02.010
KeyWords:Volumetric modulated arc therapy  Lung cancer  Planning system
FundProject:
Author NameAffiliationE-mail
Yang Jinlei Department of Radiation Oncology, Jilin Cancer Hospital, Changchun 130012, China  
Liu Boyu Department of Radiation Oncology, Jilin Cancer Hospital, Changchun 130012, China  
Lu Shuangchen 吉林大学附属第二医院放疗科  
Bu Mingwei Department of Radiation Oncology, Jilin Cancer Hospital, Changchun 130012, China zslywh@163.com 
Hits: 2810
Download times: 2763
Abstract::
      Objective To compare the performances of Monaco and Pinnacle treatment planning systems (TPS) for volumetric modulated arc therapy (VMAT) optimization regarding lung cancer. The TPS was compared in terms of dose distributions, treatment delivery parameters and quality control results.Methods For 20 patients, including 10 cases of left lung cancer and 10 cases of right lung cancer, two VMAT plans were generated for each case: one with Monaco 3.0 TPS and the other with Pinnacle 9.2 TPS. Two plans were compared among plan dosimetrie distribution. conformity index and homogeneity index of the targets, the average dose, maximum dose, minimum dose and interested dose volume histograms of organs at risk(OAR, and delivery time and MUs of the therapy plans, the accuracy of treatment plans dose verification.Results Monaco provided better PTV coverage than Pinnacle (t=5.927-12.034, P<0.05) except Dmin of PTV. Monaco had a worse sparing effect on lung than Pinnacle (t=3.545-7.485, P<0.05) except V10 of diseased side lung and V5 of total lung. In addition, Monaco had a better sparing effect on heart(t=2.836-4.011, P<0.05). Monaco had fewer delivery time(t=9.780, P<0.05)and MUs(t=5.304, P<0.05)of the therapy plans, and the pass rate of QA was better than Pinnacle(t=4.937, P<0.05).Conclusions For lung cancer patients, VMAT treatment plans obtained with Monaco and Pinnacle could offer clinically acceptable dose distributions. Pinnacle might have a better sparing effect on lung and fewer delivery time and MUs of the therapy plans. Monaco had a better PTV coverage and heart sparing. Moreover,Monaco had better accuracy of treatment plans dose verification.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9077177  On-line:0

v
Scan QR Code
&et=8C10CEA789A144B89353665E436587B85ED6378D7CE096B722C6B05932622063F45C9715AB7C344EC1FC614B4CE01779A5A5A5B6CA377076B256E893517668E09783B5FB3FDEDD90230CEDD9C8FFAC918FC6F930ED86287299F641A73ADEABA1064C85CDB03E3FD3FA295BF938F25D6F0CB72805BE7FD6D5777802D9DE78FF36&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=8A8C7B437C22286A1E45788D3418430D&vid=&iid=0B39A22176CE99FB&sid=EFD65B51496FB200&eid=B62E0EEFE746E568&fileno=20150210&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="8A8C7B437C22286A1E45788D3418430D";