Ma Yongzhong,Yang Bo,Feng Zechen,Wang Hongfang,Wan Ling.Study on dosimetric and radiation safety performance parameters of helical tomotherapy unit[J].Chinese Journal of Radiological Medicine and Protection,2015,35(1):62-68
Study on dosimetric and radiation safety performance parameters of helical tomotherapy unit
Received:October 21, 2014  
DOI:10.3760/cma.j.issn.0254-5098.2015.01.012
KeyWords:Tomotherapy unit  Helical tomotherapy  Dosimetric and radiation safety performance parameter  Quality control
FundProject:
Author NameAffiliationE-mail
Ma Yongzhong Center for Disease Control and Prevention, Beijing 100013 xmayg@sina.com 
Yang Bo Peking Union Medical College Hospital  
Feng Zechen Center for Disease Control and Prevention, Beijing 100013  
Wang Hongfang Center for Disease Control and Prevention, Beijing 100013  
Wan Ling Center for Disease Control and Prevention, Beijing 100013  
Hits: 2621
Download times: 2455
Abstract::
      Objective To understand the functional status and radiation protection requirements of helical tomotherapy (TOMO) unit when it is applied in clinic treatment so as to promote the construction of TOMO quality control system. Methods A helical tomotherapy unit with a type of tomotherapy Hi-Art was employed as the radiotherapy unit. A series of tests and analyses on dosimetric and radiation safety performance parameters of the TOMO unit were conducted with the use of a A1SL ionization chamber, a set of solid water equivalent phantoms and other test equipments. The performance parameters were compared with that of a conventional 6 MV accelerator with a type of ARTISTE. Results The deviations of static and dynamic output dose were -1.51% and 1.18%, respectively. The values of PDD10/PDD1.5 and PDD20/PDD1.5 were 0.608 and 0.318 within the range of X-ray quality parameters calibration. Under the different field conditions of 40 cm×5 cm, 40 cm×2.5 cm and 40 cm×1 cm,the stability of transverse/longitudinal dose distribution curve was no more than 1.20% and the offset distance of Jaw width was smaller than 0.5 mm. Radiation leakage ratio of the treatment rotating center from the multileaf collimator (MLC) was 0.20%. No matter the average or the maximum value of the leakage radiation from couch plane was less than 0.02%. Conclusions The performances and testing requirements of TOMO unit are different from a conventional electron linear accelerator, the quality control index of TOMO unit should be set seperately and tested strictly in terms of the specific requirements of the unit by complying with the QC programme. The reliable quality control system should be established and effectively implemented to serve the radiotherapy quality and radiation safety effectively in the tomotherapy.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9060091  On-line:0

v
Scan QR Code
&et=08EE5A5866ACE440A03C69E370D98EA7D233AFD49B05F6EB41A6EE8C08566EB357606BDA5C4DC9000910ED4C52D690F0E44D72FCC5320AE7EA87250EC81BFE81D47A5EA14FA4DD27623919B7803CCA46814BB1DDE92ED8B374D003F0979E8482&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=FFD10F7019FAA9EC&aid=35F5A1BD26637338609CD41DFE823631&vid=&iid=CA4FD0336C81A37A&sid=95D537AC89B28832&eid=68D88C2FCF9C3098&fileno=20150112&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="FFD10F7019FAA9EC"; var my_aid="35F5A1BD26637338609CD41DFE823631";