Du Changyu,Liu Yijun,Wei Wei,et al.Application value of auto-prescription technique combined with iterative reconstruction algorithm in low-dose CT pulmonary angiography[J].Chinese Journal of Radiological Medicine and Protection,2025,45(7):685-691
Application value of auto-prescription technique combined with iterative reconstruction algorithm in low-dose CT pulmonary angiography
Received:July 25, 2024  
DOI:10.3760/cma.j.cn112271-20240725-00280
KeyWords:Pulmonary CT angiography  Pulmonary embolism  Low dose  Auto-prescription
FundProject:
Author NameAffiliationE-mail
Du Changyu Department of Radiology, First Affiliated Hospital of Dalian Medical University, Dalian 116011, China  
Liu Yijun Department of Radiology, First Affiliated Hospital of Dalian Medical University, Dalian 116011, China yijunliu1965@126.com 
Wei Wei Department of Radiology, First Affiliated Hospital of Dalian Medical University, Dalian 116011, China  
Hu Mengting Department of Radiology, First Affiliated Hospital of Dalian Medical University, Dalian 116011, China  
Zhang Jingyi Department of Radiology, First Affiliated Hospital of Dalian Medical University, Dalian 116011, China  
Cheng Qiye Department of Radiology, First Affiliated Hospital of Dalian Medical University, Dalian 116011, China  
He Jian Department of Radiology, First Affiliated Hospital of Dalian Medical University, Dalian 116011, China  
Chen Anliang Department of Radiology, First Affiliated Hospital of Dalian Medical University, Dalian 116011, China  
Hits: 4179
Download times: 995
Abstract::
      Objective To explore the application value of the double-low technique of auto-prescription technique combined with iterative reconstruction algorithm in CT pulmonary angiography (CTPA). Methods A total of 86 patients who were clinically suspected of having pulmonary embolism and underwent CTPA examination in the First Affiliated Hospital of Dalian Medical University were prospectively collected and randomly assigned to a control group (n = 45) and an observation group (n = 41) according to the random number table method. In the control group, a tube voltage of 120 kVp was used with a standard iodine contrast agent dose of 60 ml, and images were reconstructed using the 40% adaptive statistical iterative reconstruction algorithm (ASIR-V). In the observation group, the tube voltage was set by auto-prescription technique, and 0.4 ml/kg of personalized low iodine contrast agent was used. Images were reconstructed with 40%, 60%, and 80% ASIR-V, respectively, and designated as observation 1, observation 2, and observation 3 respectively. The volume CT dose index (CTDIvol), dose-length product (DLP), and effective dose (E) were recorded and compared among the four groups. The CT values and standard deviation (SD) of the main pulmonary artery, left and right pulmonary arteries, as well as the left and right pulmonary lobe arteries were measured, and the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of these arteries were calculated. Additionally, the SD value at the contrast medium concentration in the superior vena cava was measured, and the artifact index (AI) was subsequently calculated. Two observers independently assessed the visibility of the pulmonary arteries, image noise, and sclerosis artifacts in the superior vena cava using a blinded method. Results The E in the observation group was 3.28 (2.08, 3.93) mSv, which was significantly lower than that in the control group [5.03 (4.86, 5.20)] mSv, and the difference was statistically significant (Z = 174.00, P < 0.05). The contrast agent dosage in the observation group was 28 (25, 30) ml, which was lower than that in the control group (60 ml), and the difference was statistically significant (Z = 0, P < 0.05). The CT values for the main pulmonary artery and the left and right pulmonary lobe arteries in the observation group were higher than those in the control group, and the differences were all statistically significant (t = -3.65 to -3.89, P < 0.05). The SNR and CNR of the observation groups 2 and 3 were greater than those of the control group (t = -9.20 to -2.98, P < 0.05). The consistency of subjective evaluations between the two observers was good (Kappa = 0.729 - 0.879, P < 0.05). There was no statistically significant difference in the subjective score of pulmonary artery visibility between the control and observation group (P > 0.05). The subjective scores for image noise in observation group 2 and group 3 were higher than those in the control group (U =598.50, 654.00, P < 0.05). The presence of artifacts due to sclerosis in the superior vena cava was significantly lower in the observation group compared to the control group (χ2 = 46.09, P < 0.001). Conclusions The combination of auto-prescription technique with ASIR-V reconstruction algorithm and low contrast agent imaging protocol can reduce the radiation dose and contrast agent dose without compromising image quality, and enable personalized double low CTPA imaging.Clinical trial registration Chinese Clinical Trial Registry, YJ-KY-2022-57
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:17875073  On-line:0

v
Scan QR Code
&et=DCA5CABA83612A94F0F16D09AE51DD0E62168134C0ACDB6C87F4C2BF59C64DF924F614CE41EE1E7BBEC1342900D93B32441DA04FC0041FE2063D22CC5D892A3418F51775E074E63478DE45DFFABC46CC20FE4C11B358C32ED53B5873A9C3132646BD0EDA83FF4D4924B59DA4D5DF6074DDB81736BE2DDBFD9A8403F765583FE4E642E81C407CF289&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=E6A891D2134F30C0&aid=6F877BFB6D17052E0539B73AB1A15DDF&vid=&iid=DF92D298D3FF1E6E&sid=AF14A8B15FB15A64&eid=09F7C8D609E885AE&fileno=20250710&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="E6A891D2134F30C0"; var my_aid="6F877BFB6D17052E0539B73AB1A15DDF";