Zhang Bingjie,Wang Jianwei,Mao Yuxuan,Jia Tianhe,Wang Xuan,Liu Tongzhen,Xu hui,Yue Baorong.Investigation and analysis of the testing results of interventional radiology equipment causing radiation skin injury[J].Chinese Journal of Radiological Medicine and Protection,2024,44(4):292-297
Investigation and analysis of the testing results of interventional radiology equipment causing radiation skin injury
Received:March 16, 2023  
DOI:10.3760/cma.j.cn112271-20230316-00079
KeyWords:Radiation skin injury  Interventional Radiology  Radiation Protection Testing
FundProject:
Author NameAffiliationE-mail
Zhang Bingjie Prevention and Treatment Center for Occupational Disease of Xinxiang City, Xinxiang 453003, China  
Wang Jianwei Prevention and Treatment Center for Occupational Disease of Xinxiang City, Xinxiang 453003, China  
Mao Yuxuan Prevention and Treatment Center for Occupational Disease of Xinxiang City, Xinxiang 453003, China  
Jia Tianhe Prevention and Treatment Center for Occupational Disease of Xinxiang City, Xinxiang 453003, China  
Wang Xuan Prevention and Treatment Center for Occupational Disease of Xinxiang City, Xinxiang 453003, China  
Liu Tongzhen Prevention and Treatment Center for Occupational Disease of Xinxiang City, Xinxiang 453003, China  
Xu hui Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, WHO Collaborating Centre for Radiation and Health, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China xuhui@nirp.chinacdc.cn 
Yue Baorong Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, WHO Collaborating Centre for Radiation and Health, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Hits: 219
Download times: 75
Abstract::
      Objective To provide suggestions for the optimization of radiation protection during interventional radiology procedures on the basis of investigate the quality control testing result of an interventional radiology equipment that has caused radiation skin damage. Method Five quality control testing reports on an interventional radiology device that has caused radiation skin injury to a patient in the past three years were selected to compare the differences between the testing results and analyze the existing problems. Results For the program of " Typical value of entrance surface air kerma in fluoroscopy mode", the testing result from the three institutions were in the range of 6.08-24.89 mGy/min, which met the requirements of relevant standards. The testing result of both the entrance surface air kerma and the ambient dose equivalent rate on the area of the operators differs greatly dependent on different exposure modes (normal fluoroscopic dose rates mode, high fluoroscopic dose rates mode, and cine mode) and different frame rates. The interventional operators for the device were lack of understanding of the influence on exposure modes, and patient doses were not recorded after an interventional surgery. Conclusions The patient dose can be significantly reduced by adjusting an interventional radiology equipment. It is recommended to increase the testing of reference point cumulative air kerma or dose-area product in future standard revisions and strengthen professional training for interventional operators and technicians, so that they can fully understand the differences in the effect of exposure mode of a interventional radiology equipment on patient and operator dose.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:8995015  On-line:0

v
Scan QR Code
&et=9A2B3ECE8B55D590AE99BF1AE9328FD22387E875AEADCACCC6336F3EA297C1A63638C72DC8C23E70BD29E423287E3AF03C68A3331ED4C9D73CCDA3300076C6406936AFB63834BEC1F8A2D8DC6A19A9AEA48A920B7ED37630CDB96D4D8299B7F9CA4EAC10F318BCE178D4E0F008CB8A9AA5FD88A1369D0B21&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=B80136CCD8DCBAA1&aid=&vid=&iid=E158A972A605785F&sid=8BB50A069C48D50B&eid=E42CAFB11D4BE21A&fileno=20240407&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="B80136CCD8DCBAA1"; var my_aid="";