Meng Xiangyin,Yang Bo,Zhu Qizhen,Liang Yongguang,Zhou Zhen,Qiu Jie.Application of stereotactic body radiotherapy planning verification phantom in HyperArc technique for brain metastases[J].Chinese Journal of Radiological Medicine and Protection,2023,43(12):991-996
Application of stereotactic body radiotherapy planning verification phantom in HyperArc technique for brain metastases
Received:March 17, 2023  
DOI:10.3760/cma.j.cn112271-20230317-00082
KeyWords:StereoPhan phantom  SRS MapCHECK semiconductor matrix  HyperArc  Dose verification
FundProject:国家重点研发计划资助(2022YFC2404606);中央高水平医院临床科研业务费资助项目(2022-PUMCH-B-116)
Author NameAffiliationE-mail
Meng Xiangyin Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100730, China  
Yang Bo Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100730, China  
Zhu Qizhen Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100730, China  
Liang Yongguang Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100730, China  
Zhou Zhen Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100730, China  
Qiu Jie Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100730, China qiujie@pumch.cn 
Hits: 745
Download times: 268
Abstract::
      Objective To explore the feasibility of StereoPhan (SP) phantoms and SRS MapCHECK (SMC) semiconductor matrices in the dose verification of HyperArc (HA) plans for patients with brain metastases (BM).Methods A total of 16 BM patients who received HA radiotherapy in the Peking Union Medical College Hospital were enrolled in this study. The ion chamber and the SMC semiconductor matrix were inserted into SP phantoms, respectively. The point and planar doses in HA verification plans were measured and compared with the calculated data of the treatment planning system (TPS). The criteria for planar dose γ analysis were set at 2 mm/3%, 2 mm/2%, 1 mm/3%, and 1 mm/2%.Results The mean deviation of point doses in 16 patients was 1.33%±0.84%. Based on the above criteria, the γ pass rates of planar doses were 99.72%±0.46%, 98.93%±1.32%, 98.85%±1.79%, and 97.13%±3.19%, respectively.Conclusions SP phantoms and SMC semiconductor matrices are applicable to the dose verification of HA plans for BM patients. The analytical criteria of 3% and 1 mm/2% can be used for verifying point and planar doses, respectively.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9077670  On-line:0

v
Scan QR Code
&et=DCA5CABA83612A949DB8F12614CC4D3FA811537550F4CA186C8A005C245D55709A4801752D8E5EC58AECD71D6A31C734110896E33121877AF977F2C7B3D2FB550FAF83D19103DD0D80ACEC0E70C4AE8CAAE3348498B7F7A5409EDBB7258111FE3533524334590517082235519BE38870DF6C17E17C8EF3D9&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=BA1E75DF0B7E0EB2&aid=6AED79CCC03FD48682028FFA5BB17B92&vid=&iid=59906B3B2830C2C5&sid=243DE6042D93E88B&eid=112A5CAF55F27887&fileno=20231208&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="BA1E75DF0B7E0EB2"; var my_aid="6AED79CCC03FD48682028FFA5BB17B92";