Ma Xiangjie,Li Mingsheng,Sun Quanfu,et al.Radiation protection testing and radiation shielding performance evaluation for a novel self-shielding radiotherapy system[J].Chinese Journal of Radiological Medicine and Protection,2023,43(11):906-912
Radiation protection testing and radiation shielding performance evaluation for a novel self-shielding radiotherapy system
Received:August 31, 2023  
DOI:10.3760/cma.j.cn112271-20230831-00070
KeyWords:Radiotherapy  Self-shielding  Radiation protection  Dosimetry  ZAP-X
FundProject:
Author NameAffiliationE-mail
Ma Xiangjie Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Li Mingsheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Sun Quanfu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Wang Chengcheng Department of Radiation Oncology, First Medical Center of PLA General Hospital, Beijing 100853, China  
Wei Xiao Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Cheng Jinsheng Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China chengjs3393@163.com 
Hits: 2399
Download times: 1084
Abstract::
      Objective To evaluate the radiation shielding performance for a novel self-shielded ZAP-X radiotherapy system used for intra-cranial and neck treatments. The present evaluation was performed according to the relevant Chinese national standards and the clinical placement of the radiotherapy system in an unshielded treatment room.Methods The radiation source of the ZAP-X was a 3 MV linear accelerator. A total of 33 detection sites were selected surrounding the self-shielded system at 1.3, 2.3 and 3.3 m away from the periphery of the equipment. The maximum ambient dose equivalent rate in each irradiation condition was measured accordingly. A commonly used clinical treatment plan was selected to simulate the treatment process. During the delivery of this treatment plan, the cumulative doses of these 33 sites were measured, separately. The applicable current radiation protection standard for radiotherapy in China was chosen to evaluate the radiation shielding performance of the system.ResultsAccording to the measurement result of the ambient dose equivalent rates along the aforementioned perimeter lines, a suggestion was put forward to redefine the existing 1 m controlled area by determining the distance at which the instantaneous dose rate of 10 μSv/h will not be exceeded. This is to meet the requirements of the Chinese standard GBZ 121-2020.Conclusions According to the existing Chinese national radiation protection standards, the self-shielded radiotherapy system in the unshielded treatment room has the clinical applicability in China. But for such a novel self-shielded system, the corresponding performance testing and radiation protection standards shall be formulated.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12034578  On-line:0

v
Scan QR Code
&et=15C444C2D45CD59A83230A5334377401D5B410FFD645891739997963B28231367DD7023A4922639DB869734B35E5AC259531AB8A49882884A00FB2C550104C340CA2F24B414067F9924C5E3576616E0A812B5DBCBD299CD58257AA6CAE5DD7AF59E1793D07E85DD674DB203325F3F6130108B01B090F3DF810CB7A372AAD817C&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=BA1E75DF0B7E0EB2&aid=6680D355800CC8EE941A4BD21A85F254&vid=&iid=708DD6B15D2464E8&sid=0FCE62EC9DC00F5E&eid=F434A3C2A19884E7&fileno=20231109&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="BA1E75DF0B7E0EB2"; var my_aid="6680D355800CC8EE941A4BD21A85F254";