Wang Hongkai,Li Minghui,Niu Chuanmeng,Song Yixin,Han Dongsheng,Men Kuo,Dai Jianrong.Challenge of shielding design for FLASH radiotherapy[J].Chinese Journal of Radiological Medicine and Protection,2023,43(8):653-656 |
Challenge of shielding design for FLASH radiotherapy |
Received:November 07, 2022 |
DOI:10.3760/cma.j.cn112271-20221107-00435 |
KeyWords:Radiation protection Dose limit Instantaneous dose rate Standards |
FundProject:北京市自然科学基金(1232033);国家自然科学基金(12205374);中国癌症基金会北京希望马拉松专项基金(LC2021B23) |
Author Name | Affiliation | E-mail | Wang Hongkai | Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China | | Li Minghui | Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China | | Niu Chuanmeng | Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China | | Song Yixin | Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China | | Han Dongsheng | Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China | | Men Kuo | Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China | | Dai Jianrong | Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China | dai_jianrong@163.com |
|
Hits: 840 |
Download times: 374 |
Abstract:: |
Compared with conventional radiotherapy, FLASH radiotherapy has advantages in protecting normal tissues, while the dose rate is increased by more than 100 times. If the shielding design of the treatment room is carried out according to the existing standard, the thickness and cost of the shielding wall will be significantly increased, or even hardly to meet the requirement of the standards, resultsing in the failure of the application of FLASH radiotherapy. By investigating the domestic and foreign standards and literature, this paper analyzes the challenges brought by FLASH radiotherapy technology to the shielding design of radiotherapy treatment room in China. Dose rate control standards adopted by different countries in the shielding design are emphatically compared as well. In several countries, the average dose rate under the actual treatment conditions was considered in the shielding design. In China, the method of instantaneous dose rate taking acount of occupancy factor is adopted. However, if FLASH radiotherapy technology is applied, the requirement of instantaneous dose rate will be difficult to meet. In order to improve the high dose rate radiotherapy technology such as FLASH radiotherapy, the revision of the existing standards is advised if the authorized limits are not changed. To use the average dose rate limit within a certain period of time for control, or to raise the control standard in the case of flash radiotherapy, are also avaliable. |
HTML View Full Text View/Add Comment Download reader |