Fu Qiang,Wang Lu,Xi Yue,Sun Liang,Ji Shengrong,Ren Zhonghao,Wang Jia,Liu Bing.Study on the effect of personal radiological protective equipment used in diagnostic radiology[J].Chinese Journal of Radiological Medicine and Protection,2023,43(6):462-468
Study on the effect of personal radiological protective equipment used in diagnostic radiology
Received:October 21, 2022  
DOI:10.3760/cma.j.cn112271-20221021-00415
KeyWords:Personal protective device  Ambient dose equivalent  Concern point  Administrative law enforcement
FundProject:
Author NameAffiliationE-mail
Fu Qiang Radiation Protection Management Section, Qingdao Hospital, University of Health and Rehabilitation Sciences (Qingdao Municipal Hospital), Qingdao 266011, China  
Wang Lu Radiation Protection Management Section, Qingdao Hospital, University of Health and Rehabilitation Sciences (Qingdao Municipal Hospital), Qingdao 266011, China  
Xi Yue Radiation Protection Management Section, Qingdao Hospital, University of Health and Rehabilitation Sciences (Qingdao Municipal Hospital), Qingdao 266011, China  
Sun Liang School of Radiology and Protection, Soochow University, Suzhou 215127, China  
Ji Shengrong Qingdao Huanwan Testing and Evaluation Co., LTD, Qingdao 266100, China  
Ren Zhonghao Qingdao Huanwan Testing and Evaluation Co., LTD, Qingdao 266100, China  
Wang Jia Qingdao Huanwan Testing and Evaluation Co., LTD, Qingdao 266100, China  
Liu Bing Law Enforcement Supervision Bureau, Shandong Provincial Health Commission, Jinan 250002, China sdjnliubing@126.com 
Hits: 1306
Download times: 354
Abstract::
      Objective To study the actual effect of the use of personal protective equipment of the examined individuals, and provide reference and basis for the correct use of personal protective equipment and the radiological health administrative law enforcement.Methods From February to June 2022, the imaging department of Qingdao Municipal Hospital selected 170 patients who underwent X-ray imaging examination (oral panoramic, dental radiography, DR photography, CT scanning), including 25 with oral panoramic and dental radiography, 60 with CT scanning and 60 with DR imaging. The thermoluminescent dosimeter was used to detect the ambient dose equivalent at the point of concern for 170 examined individuals who have used personal protective equipment to cover their sensitive parts, and to analyze the data detected at the same point as above when routinely using the same equipment.Results There was a statistically significant difference in the dose equivalent at the same points inside and outside the lead neckband (t=-2.23, P<0.05). There was no statistically significant difference in the dose equivalent at the same point inside and outside the lead collar during dental radiography (P>0.05). During DR photography (chest PA, lateral and lumbar AP), the examined individuals were wearing lead aprons. Among them, there was a statistically significant difference in the dose equivalent at the same points inside and outside the lead aprons of children's chest PA and adults' chest PA (U=10.00,19.00,P<0.05). There was no statistically significant difference in the dose equivalent at the same points inside and outside the lead aprons of adult's chest PA and lumbar AP (P>0.05). When performing CT scan (chest or upper abdomen), there was a statistically significant difference in the dose equivalent at the same points of wrapped lead aprons(U=878.50,11.00,P<0.05).Conclusions The correct use of personal protective equipment is a complex technical problem. It is very important to fully and accurately understand the optimization principle of radiation protection and correctly use personal protective equipment for the examined individuals. The administrative punishment of radiation health on the use of personal protective equipment of the examined individuals should be cautious.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9059502  On-line:0

v
Scan QR Code
&et=08EE5A5866ACE440F64F858F98457E477C211DD37587FE959D623819A62FE6BEEB2B7E77C09CAD3579C8BE4623C4899FEC84E59BC020769C158BD4F9487CC51C4E39304D68E4DC8B1A604BE3C2B91D8D18368C7841B564C6305F811C24497674&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=BA1E75DF0B7E0EB2&aid=7F24A11D2A9724EBB3C4E89B87BC2B08&vid=&iid=B31275AF3241DB2D&sid=8C5DE51F0A009A0F&eid=3FC4D669D19FF0C6&fileno=20230610&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="BA1E75DF0B7E0EB2"; var my_aid="7F24A11D2A9724EBB3C4E89B87BC2B08";