Zhang Daguang,Zhu Zixi,Sun Jiana,Tian Yuan,Wang Wei.Radiation shielding for the intraoperative radiotherapy using low energy X-rays[J].Chinese Journal of Radiological Medicine and Protection,2022,42(8):629-635
Radiation shielding for the intraoperative radiotherapy using low energy X-rays
Received:March 16, 2022  
DOI:10.3760/cma.j.cn112271-20220316-00103
KeyWords:Intraoperative radiotherapy  Low energy X-ray radiotherapy unit  Radiation protection  Shielding design
FundProject:国家自然科学基金(81872472);MDAnderson癌症中心全球学术项目(4-1-97)
Author NameAffiliationE-mail
Zhang Daguang Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital, National Clinical Research Center for Cancer, Tianjin's Clinical Research Center for Cancer, Key Laboratory of Cancer Prevention and Therapy, Tianjin 300060, China  
Zhu Zixi Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital, National Clinical Research Center for Cancer, Tianjin's Clinical Research Center for Cancer, Key Laboratory of Cancer Prevention and Therapy, Tianjin 300060, China  
Sun Jiana Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital, National Clinical Research Center for Cancer, Tianjin's Clinical Research Center for Cancer, Key Laboratory of Cancer Prevention and Therapy, Tianjin 300060, China  
Tian Yuan Department of Radiation Oncology, Cancer Hospital, Chinese Academy of Medical Sciences (CAMS) and Peking Union Medical College (PUMC), National Cancer Center, National Clinical Research Center for Cancer, Beijing 100021, China  
Wang Wei Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital, National Clinical Research Center for Cancer, Tianjin's Clinical Research Center for Cancer, Key Laboratory of Cancer Prevention and Therapy, Tianjin 300060, China weiwang_2@126.com 
Hits: 1665
Download times: 635
Abstract::
      Objective To assess the shielding requirements of low energy X-ray intraoperative radiotherapy room under the domestic and foreign standards and guidelines, to measure the sured transmission factors for a range of shielding materials, the ambient dose equivalent rate around concerned positions and the shielding effect of protective devices, so as to provide references for shielding design of such radiotherapy units and applications of radiological protection devices.Methods The required shielding thicknesses for a treatment room with INTRABEAM intraoperative radiotherapy system were calculated under the Chinese national standard GBZ 121, IPEM report No. 75 and NCRP report No. 151, respectively. The transmission factors for a range of shielding materials including solid water slab, shielding sheet and shielding apron were measured. Moreover, the ambient dose equivalent rates were measured under the simulated working conditions and the shielding effectiveness of a lead screen was evaluated.Results The required lead thicknesses calculated under different standards and guidelines were less than 0.6 mm for all the concerned points, with the differences at sub-millimeter level. The low energy X-rays generated by this equipment attenuated rapidly in the shielding materials. The measured transmission factors of 0.05 mm lead equivalent shielding sheet and 0.25 mm lead equivalent shielding apron were 0.068 and 0.0038, respectively. When the radiation was delivered using spherical applicator without any attenuation, the ambient dose equivalent rates at 1 m and 2 m from the X-ray source were 10.7 and 2.6 mSv/h, respectively. The corresponding measurement values decreased to 3.8 and 0.9 μSv/h, respectively, when the spherical applicator was inserted into a small water tank. Meanwhile, the ambient dose equivalent rate at 2 m was reduced to the background level when using protective screen.Conclusions The shielding requirements for a low energy X-ray intraoperative radiotherapy facility are minimal, with low effective energy of X-rays generated by this equipment, but the dose rate close to the unshielded radiation source is high. The shielding scheme of treatment room should be optimized in design and the protective device should be used in a reasonable way.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9063336  On-line:0

v
Scan QR Code
&et=15C444C2D45CD59A79A132A3D94A4F3DAD7E1583ED6528D646FCE6BF8C7C7682CCE727AA0575DAB60603CFB26D0B59BA1D04B9A1306AC199F9318337469F56E8B67B5743C1E6CA0F3072D0478DEBC28F&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=885CEFEC57DA488F&aid=68F7EBB7E610720F1581C7CC84B1B7CD&vid=&iid=5D311CA918CA9A03&sid=E4EC39E73004B593&eid=20C9FB8C7B4A22AD&fileno=20220810&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="885CEFEC57DA488F"; var my_aid="68F7EBB7E610720F1581C7CC84B1B7CD";