Yang Dewu,Sui Yan,Liu Kang,Niu Yantao.Effects of organ dose modulation technique on image quality and radiation dose in sinus CT scanning[J].Chinese Journal of Radiological Medicine and Protection,2022,42(3):230-234
Effects of organ dose modulation technique on image quality and radiation dose in sinus CT scanning
Received:October 10, 2021  
DOI:10.3760/cma.j.cn112271-20211010-00407
KeyWords:Organ dose modulation  Sinus  Eye lens  Radiation protection  Image quality
FundProject:
Author NameAffiliationE-mail
Yang Dewu Department of Medical Technique, Beijing Health Vocational College, Beijing 102433, China  
Sui Yan Department of Radiology, Fuxing Hospital, Capital Medical University, Beijing 100038, China  
Liu Kang Department of Radiology, Fuxing Hospital, Capital Medical University, Beijing 100038, China  
Niu Yantao Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China ytniu163@163.com 
Hits: 1991
Download times: 547
Abstract::
      Objective To investigate the effects of organ dose modulation (ODM) on image quality and radiation dose to lens during sinus CT scanning. Methods The GE Revolution EVO CT was used to scan the anthropomorphic head with ODM turned off as the control group and ODM turned on as the observation group. With different parameter combinations of tube voltage (140, 120, 100 kV), noise index (N17, N18), pitch (0.531, 0.969 at 20 cm collimation width, 0.516, 0.984 at 40 mm collination width), 24 series of images were acquired respectively. The tube current (mA) valuesat the anterior (A), left (L), posterior (P) and right (R) positions, volume CT dose index (CTDIvol) and dose length product (DLP) were recorded. Average CT values (AV), standard deviation (SD), signal-to-noise ratio (SNR), contrast to noise ratio (CNR) and image quality factor (FOM) of ROI were calculated. Double blind method was used to evaluate the edge sharpness, soft tissue level, noise and artifact on a subjective scale of 3 points. Results For the combinations of different scanning parameters, the radiation dose of the observation group was lower than that of the control group. The tube current attheposition A for the observation group was lower than that of the control group, and the difference was statistically significant (t=2.28, P<0.05). The differences oftube currentsatposotions of L, P and R,CTDIvol and DLP values in twogroupswere not statistically significant (P>0.05). There were no significant differences in SD, SNR, CNR and FOM between the two groups of images (P>0.05). There were no significant differences in the subjective scores of edge sharpness, soft tissue gradation, noise and artifacts between the two groups (P>0.05). ODM technique can effectively reduce the radiation dose of eye lens without decompromising the image quality, which could meet the requirements of diagnosis, and accord with the principle of radiological protection optimization.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9064865  On-line:0

v
Scan QR Code
&et=970261AF74811A5D32E699EC4F46CF537A15C51331F8B15CB9BCA30AD11EDE160EABF0A9DEC6694FD0D586F5C3C029A3C12AAE5A0A965AA58E013816BBD40448224475F5DE082964D5B6A869561AC7A60284BD9BF3460E24CE4C0C1A367F7C3372F1AE6A7452AF00&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=885CEFEC57DA488F&aid=A3BC2FF9C4CC1A9810E9D5FCEA631171&vid=&iid=38B194292C032A66&sid=D6354F61445E9456&eid=D9AE183D3F5C3C75&fileno=20220313&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="885CEFEC57DA488F"; var my_aid="A3BC2FF9C4CC1A9810E9D5FCEA631171";