Qian Yuxin,Yin Liangliang,Xie Yuhan,et al.Study on source thickness and counting efficiency calibration for determining gross alpha activity in drinking water[J].Chinese Journal of Radiological Medicine and Protection,2021,41(12):937-940
Study on source thickness and counting efficiency calibration for determining gross alpha activity in drinking water
Received:July 05, 2021  
DOI:10.3760/cma.j.issn.0254-5098.2021.12.010
KeyWords:Drinking water  Radioactivity  Gross α  Mass thickness  Counting efficiency
FundProject:国家科技基础性工作专项(2014YF211000)
Author NameAffiliationE-mail
Qian Yuxin Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Yin Liangliang Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Xie Yuhan Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Ji Yanqin Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China jiyanqin@nirp.chinacdc.cn 
Hits: 3217
Download times: 1689
Abstract::
      Objective To study the influence of source thickness and counting efficiency calibration on the measurement of gross alpha activity in water. Methods 241Am and natural uranium reference materials were spiked in drinking water to prepare source on a planchet with different thickness, for counting alpha activity on the planchet. Results The effective thickness measured by spiking 241Am or uranium standard solution in water sample was consistent with the empirical value of 4 mg/cm2. The alpha counting rate was in a linear increase trend from 2A-5A mg/cm2 and was basically stable and no longer increase when thickness was higher than 10A mg/cm2 (A was area of planchet). The result calculated by effective thickness method and thick source method were in good agreement when thickness was 10A mg/cm2. Conclusions In order to reduce the deviation of gross alpha counting rate caused by the source thickness and counting efficiency calibration, the source thickness is recommended to be 10A mg/cm2.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12029535  On-line:0

v
Scan QR Code
&et=08EE5A5866ACE44059F4E3DC20D309EDD5F7F2F6309639210DBB6FF1449209FF8CA96921BA7506788B71E01021E3001E24F0A91666162ED06E9F6320238737E563F8DC76A68F27C6813107AE52F364808DE8DAB5F048587AEEB2B320B8860E26B12ECBE792845BC03A753D60F1D1D3E1BE30DB28575792CB&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=9475FABC7A03F4AB&aid=AA4FE43705B0D4A79DAAE6BD90FC8FB9&vid=&iid=59906B3B2830C2C5&sid=AB8B0EE7E1A96CB2&eid=2BCD2FED7562E194&fileno=20211210&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="9475FABC7A03F4AB"; var my_aid="AA4FE43705B0D4A79DAAE6BD90FC8FB9";