Liu Jianxin,Cao Shenhui,Xu Tongjiang,Wang Xu.Experimental study on relative deviation between display value and measured value of entrance surface dose (ESD) in digital radiography and influencing factors[J].Chinese Journal of Radiological Medicine and Protection,2020,40(6):484-488
Experimental study on relative deviation between display value and measured value of entrance surface dose (ESD) in digital radiography and influencing factors
Received:December 10, 2019  
DOI:10.3760/cma.j.issn.0254-5098.2020.06.013
KeyWords:Digital radiography  Entrance surface dose (ESD)  Display value  Relative deviation
FundProject:
Author NameAffiliationE-mail
Liu Jianxin Department of Radiology, Emergency General Hospital, Beijing 100028, China ljx1963@aliyun.com 
Cao Shenhui Department of Radiology, Emergency General Hospital, Beijing 100028, China  
Xu Tongjiang Department of Radiology, Emergency General Hospital, Beijing 100028, China  
Wang Xu Department of Radiology, Emergency General Hospital, Beijing 100028, China  
Hits: 1973
Download times: 1018
Abstract::
      Objective To compare the relative deviation between display value and measured value of entrance surface dose (ESD) at different tube voltages, source-to-image distances and phantom thicknesses, so as to provide the basis for the feasibility of using real-time ESD display value to evaluate the radiation dose and the suitability of exposure parameters. Methods At two different tube voltages and source-to-image distances, the exposure experiments were carried out using three polyethylene models with surface area of 30 cm×30 cm and thickness of 5 cm, 10 cm and 20 cm, respectively. After each exposure, the ESD display values on DR equipment and the measured values on radiation dose detector were recorded. The Wilcoxon rank sum test was used to compare the difference between them, with the relative deviation of ESD display value calculated using formula. Results At the same tube voltage and source-to-image distance, the ESD display values were not affected by the thickness of phantom, but the measured value changed with the thickness of the phantom. When the thickness of the phantom was 10 cm and 20 cm respectively, the ESD display value was smaller than the measured value and the difference was statistically significant (Z=-2.201,-2.203,P<0.05). The thicker mold would lead to the larger relative deviation between display value and measured value. Only when the thickness of mold was 20 cm and source-to-image distance 100 cm, the relative deviation was greater than±20%. Conclusions The ESD display value on DR equipment can be used to evaluate the skin incident dose to the examinee, with necessary account taken of the influence of their body thickness.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9076852  On-line:0

v
Scan QR Code
&et=3B5D9771435A916B26445566CEFC6D4B7DEE3B35BADA655CD6C25D7F40D8F3715BA14EF65F0830DFAA4C21B4F39A18A1FC59CACF80AE95023B506C618DA44AC3BBA9CB6FC9B3E1984A834CF3C2B8DEB164522EC936E26815660FD130842A0A11A654C58A2319E08093A0CBD668BC1206DEFDF4B02E066183003578B26A5938871FC9D6D41B1C87AE51B0EB91D980BE8018514490478955A813CDF10CC132BD19&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=0D1D160AB8016934&aid=20AAC20D0D1EE54B42D4C539C97B130A&vid=&iid=B31275AF3241DB2D&sid=D8B37A0210DB6B9D&eid=8CE1095CD639AEF4&fileno=20200613&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="0D1D160AB8016934"; var my_aid="20AAC20D0D1EE54B42D4C539C97B130A";