Yang Chunyong,Zhou Yuanyuan,Cao Xingjiang,et al.Verification of doses to PTV and OAR and 2D dose distribution in IMRT for 8 accelerators in Jiangsu province[J].Chinese Journal of Radiological Medicine and Protection,2020,40(2):129-133
Verification of doses to PTV and OAR and 2D dose distribution in IMRT for 8 accelerators in Jiangsu province
Received:April 02, 2019  
DOI:10.3760/cma.j.issn.0254-5098.2020.02.010
KeyWords:Intensity modulated radiation therapy  Radiochromic film  Planned target volume  Organ at risk  Two-dimensional dose distribution
FundProject:IAEA资助项目(CPR/17821);江苏省流行病学重点学科(ZDXKA2016008);江苏省医学创新团队(CXTDA2017029)
Author NameAffiliationE-mail
Yang Chunyong Department of Radiation Protection, Jiangsu Provincial Center for Disease Control and Prevention, Nanjing 210009, China  
Zhou Yuanyuan Department of Radiation Protection, Jiangsu Provincial Center for Disease Control and Prevention, Nanjing 210009, China  
Cao Xingjiang Department of Radiation Protection, Jiangsu Provincial Center for Disease Control and Prevention, Nanjing 210009, China  
Du Xiang Department of Radiation Protection, Jiangsu Provincial Center for Disease Control and Prevention, Nanjing 210009, China  
Luo Suming Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Xue Xian Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
He Zhijian Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Wang Jin Department of Radiation Protection, Jiangsu Provincial Center for Disease Control and Prevention, Nanjing 210009, China jinwang@jscdc.cn 
Hits: 3879
Download times: 1669
Abstract::
      Objective To validate the method for measuring the TPV and OAR doses and 2D dose distribution in IMRT through using TLD and radiochromic film. Methods Eight medical linear accelerators (Valian, Elekta, Siemens) were selected. The polystyrene phantom provided by IAEA was CT scanned and the image obtained was transferred to TPS for formulation of treatment plan, prescription of PTV and OAR doses and calculation of corresponding monitoring unit (MU), IMRT was performed on the phantom using 6 MV X-ray. Irradiated TLDs and films were measured and evaluated at the Secondary Standard Dosimetry Laboratory at the Radiation Safety Institute of Chinese Center for Disease Control and Prevention. Results According to IAEA requirement, the relative deviations between TLD-measured and TPS-planned doses were within±7.0% for the prescribed PTV and OAR doses. As measured result, the PTV values for 8 accelerators were in the range of 0.6% to 5.9%, consistent with the IAEA requirements, whereas the OAT values for 8 accelerators were within -0.6% to 7.0%, consistent the requirements. As IAEA required, the 2D dose distribution passing rate of 3 mm/3% should be higher than 90%. The film-measured and TPS-planned values for 8 accelerators were within 90.2% to 100.0%, consistent with the requirements. Conclusions TLD and radiochromic film are feasible in validating the PTV and OAR doses and the 2D dose distribution pass rate in IMRT. This method can be widely used in quality audit and internal verification in IMRT in medical institutiions on a large scale.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12034959  On-line:0

v
Scan QR Code
&et=BF7E82E5F022C752D74314D26FEFC0EABBCBDB972265694C21AB8D1CB3C05E3509A6AA757E0808C8011FD2CEA3A6340432F6A1023F605141DB5E2AC8E8A068AA983A754844B12B1DBF1AA3A92ECBC7119BC210801E6F9ED8C02CD3C1A295F9354AE1718847D5E97DDEF5E94B9C8BB251&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=0D1D160AB8016934&aid=4145DAB3660DF776A755FF28A67F12AD&vid=&iid=0B39A22176CE99FB&sid=28F8B56DB6BEE30E&eid=76B5E24D6EC46B4B&fileno=20200210&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="0D1D160AB8016934"; var my_aid="4145DAB3660DF776A755FF28A67F12AD";