Development of methodology for quality audit of doses in target area and organ at risk and two dimensional dose distribution in intensity modulated radiotherapy
Received:March 27, 2019  
DOI:10.3760/cma.j.issn.0254-5098.2020.02.009
KeyWords:Intensity modulated radiation therapy  Radiochromic film  Planned target volume  Organ at risk  Two-dimensional dose distribution
FundProject:IAEA资助项目(CPR/17821)
Author NameAffiliation
Luo Suming Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Wu Hao Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing 100036, China 
He Zhijian Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Xue Xian Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Yuan Jilong Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Hits: 3936
Download times: 1691
Abstract::
      Objective To develop the methodology for using TLD and radiochromic film to measure the planned target volume (PTV) and organ at risk (OAR) doses and 2D dose distribution in IMRT, in order to provide technical guidance on the dose quality audit in IMRT at home. Methods China has participated in the research project launched by the international multi-radiotherapy centre (IMRC). IMRT polystyrene phantom provided by IAEA was scanned by CT scanner and then the scanned images were transmitted to TPS to outline prescribed dose to PTV and to OAR. The former was limited to 400 cGy while the latter limited to 200 cGy. IMRT was implemented with the phantom irradiated using 6 MV X-ray. The irradiated TLDs and films were sent to IAEA dosimerty laboratory for measurement and calculation. Jiangsu, Sichuan, Hubei and Henan provinces were selected to engage with this study for their variety of accelerators and highly skilled physicists. The procedures used were the same as in the IMRC and the irradiated TLDs and films were required to send to external audit group for measurement and calculation. Results According to IAEA requirement, the relative deviations of the TLD-measured and TPS planned doses are within±7.0% for PTV and OAR. The China's research results at the IMRC have shown that the relative deviation of TLD-measured and TPS-planned values for the upper and lower PTV were -0.2% and 0.8%, respectively, consistent with the IAEA requirement, and the values for upper and lower OAR were -0.6% and -1.0%, respectively, consistent with the requirement. As the results have shown in four provinces, the relative deviations of the TLD-measured and TPS-planned were within 0 to 10.6% for upper and lower PTV and -0.6% to 20.9% for upper and lower OAR. According to IAEA requirement, the passing rate should be greater than 90% for 3 mm/3% for 2D dose distribution. China's result at the IMRC is 100%, being excellent. The four provinces' results have shown that 2D dose distribution pass rate of 3 mm/3% was in the range of 45.0%-100.0%. Conclusions The uses of TLD in quality audit for PTV and OAR doses and the radiochromic film in 2D dose distribution pass rate in IMRT are characterized by scientific feasibility, strong operability, easy-to-mail and data realibility. They are can be applied to quality assurance and audit in medical institutions in the country to on a large scale.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12030073  On-line:0

v
Scan QR Code
&et=A90EAC7419FB9B22C5C57F2085C69769189A7B80470D296C32127F9E67E2EF3ABE3528D82EFFF5590C06D7B4ED613A3DD5013BF8578EEFF729CB39CEF2947AA6CFB74EB3D76C5A8BCB030BD2B417166277BB49ADEF0BC17046BA6C92A7449AB88BCB863A3658106C7493CF62076B0505A89B73B37B3022D8C1EB436FD1136EAE0310F2D2DE455BD8E45A97338F395AB86D1B5C9C429D4FBAC1F4D48182F9071895F7F1C04552495D&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=0D1D160AB8016934&aid=4145DAB3660DF776B8209DA7A2EF7DFB&vid=&iid=0B39A22176CE99FB&sid=B62E0EEFE746E568&eid=1F199509C0B6C4D6&fileno=20200209&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="0D1D160AB8016934"; var my_aid="4145DAB3660DF776B8209DA7A2EF7DFB";