Li Wenyan,Zhang Guiying,Wang Zhichao,Guo Zeqin,Kang Mengxiao,Liu Chuanye,Sheng Wentao,Zeng Yue,Luo Xi,Yue Linlin.Research of TLD calibrated in different standard X-ray RQR radiation field in terms of eye lens dose Hp(3)[J].Chinese Journal of Radiological Medicine and Protection,2020,40(1):42-46 |
Research of TLD calibrated in different standard X-ray RQR radiation field in terms of eye lens dose Hp(3) |
Received:May 24, 2019 |
DOI:10.3760/cma.j.issn.0254-5098.2020.01.007 |
KeyWords:Eye lens RQR field Hp(3) Thermoluminescent dosimeter(TLD) |
FundProject: |
Author Name | Affiliation | E-mail | Li Wenyan | Radiological Protection Measurement Office, Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry, Beijing 100093, China | | Zhang Guiying | Radiological Protection Measurement Office, Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry, Beijing 100093, China | | Wang Zhichao | Radiological Protection Measurement Office, Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry, Beijing 100093, China | | Guo Zeqin | Radiological Protection Measurement Office, Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry, Beijing 100093, China | | Kang Mengxiao | Radiological Protection Measurement Office, Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry, Beijing 100093, China | | Liu Chuanye | Radiological Protection Measurement Office, Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry, Beijing 100093, China | | Sheng Wentao | Radiological Protection Measurement Office, Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry, Beijing 100093, China | | Zeng Yue | Radiological Protection Measurement Office, Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry, Beijing 100093, China | | Luo Xi | Radiological Protection Measurement Office, Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry, Beijing 100093, China | | Yue Linlin | Huaneng Shandong Shidaowan Nuclear Power Limited Company, Rongcheng 264300, China | yuelinlin83@126.com |
|
Hits: 2190 |
Download times: 1160 |
Abstract:: |
Objective To compare the calibration result of standard X-ray RQR radiation field between SSDL (NIRP) and CEA LIST LNHB(France), and to explore the feasibility of calibrating Hp(3) in standard X-ray RQR radiation field of SSDL(NIRP). Methods Using a column model with a diameter and high of 20 cm, TLD was calibrated in SSDL (NIRP) and CEA LIST LNHB (France) to measure the personal dose equivalent eye lens dose Hp(3), X-ray RQR radiation field included RQR4(60 kV), RQR7(90 kV), RQR9(120 kV), with energy response, angle response and linear response. Results In terms of energy response, the calibration results of TLD in both SSDL (NIRP) and CEA LIST LNHB (France) were in good agreement. The difference between exposure value and response value was less than 10%. In terms of angle response, the calibration result of TLD in CEA LIST LNHB (France) was better in SSDL(NIRP). The difference between exposure value and response value in CEA LIST LNHB (France) was less than 6%, while the difference between exposure value and response value in SSDL(NIRP) was more than 10% at angle of 20°. In terms of linear response, both calibration result of SSDL (NIRP) and CEA LIST LNHB (France) were in good agreement. Conclusions The standard X-ray RQR field in SSDL (NIRP) can be used for the calibration of Hp(3). |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|