Zhang Yongxian,Niu Yantao,Liu Dandan,Li Wei,Zhang Lili,Wu Jianxing,Kang Tianliang,Guo Senlin.A phantom study of the effects of tube voltage combined with organ dose modulation on radiation dose and image quality in chest CT[J].Chinese Journal of Radiological Medicine and Protection,2019,39(7):529-533
A phantom study of the effects of tube voltage combined with organ dose modulation on radiation dose and image quality in chest CT
Received:January 11, 2019  
DOI:10.3760/cma.j.issn.0254-5098.2019.07.009
KeyWords:Organ dose modulation  Superficial radiation sensitive organ  Image quality  Radiation dose  Tube voltage
FundProject:国家自然科学基金(81872559)
Author NameAffiliationE-mail
Zhang Yongxian Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Niu Yantao Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China ytniu163@163.com 
Liu Dandan Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Li Wei Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Zhang Lili Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Wu Jianxing Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Kang Tianliang Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Guo Senlin Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Hits: 3224
Download times: 1372
Abstract::
      Objective To investigate the effects of different tube voltages on the dose of superficial radiation-sensitive organs and image quality when using organ dose modulation(ODM) in chest CT. Methods Based on clinical chest CT protocol with the sameother parameters, chest phantom was scanned using 140, 120, 100, and 80 kv (100 kV was the recommended by the CT system) without ODM (ODM off) or with ODM from the starting layer to the breast area (ODM part). A long rod ionization chamber was placed iat a fixed position in front of the right breast area. The scans were repeated for 7 times with each group of scanning parameters and dose values were measured for each scanning, the CTDIvol and breast skin dose measurements(D) were recorded. Coronal images of 5 mm thickness for the lung and soft tissue algorithms were reformatted. The images were divided into 8 parts along the z axis direction, the contrast noise ratios(CNR) for every region were measured. For CTDIvol, D, CNR for different ODM and tube voltage scanning modes, two factor non-repeat test ANOVA was performed. LSD method was used for comparison among groups. Results The CTDIvol was lowest at 80 kV, and the breast skin dose measurement was lowest at 100 kV, CTDIvol decreased in turn from140 to 80 kV (F=105.579 5, P<0.05). The breast skin dose measurement decreased in turn from140 to 100 kV, but increased instead at 80 kV. The difference was statistically significant(F=27.736, P<0.05). Compared with ODM off, the CTDIvol and D for ODM part both declined and the differences were statistically significant(F=39.732, 81.961, P<0.05). The CNRs of the lung and soft tissue images decreased at every tube voltage(F=12.809, 11.261, P<0.05).The CNRs decreased from140 to 100 kV, but there was no statistical difference(P>0.05), and the difference was significant at 80 kV(P<0.05). Compared with ODM off, the CNRs of lung and soft tissue algorithm images with ODM part decreased, withnot statistically significant differences(P>0.05). Conclusions In clinical practice, with the tube voltage not less than the recommended(100 kV), the optimal reduction of breast radiation dose can be achieved by reducing kV and using ODM on the premise of resonable image quality.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9057821  On-line:0

v
Scan QR Code
&et=5128A264A75270F483CC666099D83C610C1FBF8E3BC70ED6634E19800FF16F39A3E9175B97498E12A2DC3489BB9EB65F9C9764AD37F757061E6E6AED956C49326AEA4C74F1112BFD49425F4FC3CEC29112732D6BD33422CEC46BA133071109811FA1FB234AE921763198E94DA67F3A5A1B117EE8D54B7EB066D2F869AF94FBD0B77EB395085EFCFE&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=B6351343F4791CA3&aid=7BA62EAAEF4CF3CA4D8C3E036DAC2554&vid=&iid=DF92D298D3FF1E6E&sid=2A2AA8B7E19F0DF7&eid=FF58680609C9D068&fileno=20190709&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="B6351343F4791CA3"; var my_aid="7BA62EAAEF4CF3CA4D8C3E036DAC2554";