Yang Ke,Yu Xirong,Zhang Xianpeng,et al.Study on estimation methods of absorbed dose to eye lens of infants from CT head and neck scanning based on one-year old anthropomorphic phantom[J].Chinese Journal of Radiological Medicine and Protection,2019,39(4):297-302
Study on estimation methods of absorbed dose to eye lens of infants from CT head and neck scanning based on one-year old anthropomorphic phantom
Received:December 11, 2018  
DOI:10.3760/cma.j.issn.0254-5098.2019.04.009
KeyWords:Infants  CT  Eye lens  Absorbed dose  Estimation methods
FundProject:国家自然科学基金(11475047)
Author NameAffiliationE-mail
Yang Ke Institute for Radiological Protection and Safety, Shandong Center for Disease Control and Prevention, Jinan 250014, China  
Yu Xirong Institute for Radiological Protection and Safety, Shandong Center for Disease Control and Prevention, Jinan 250014, China  
Zhang Xianpeng Institute for Radiological Protection and Safety, Shandong Center for Disease Control and Prevention, Jinan 250014, China  
Liu Yi Institute for Radiological Protection and Safety, Shandong Center for Disease Control and Prevention, Jinan 250014, China  
Tang Bo Institute for Radiological Protection and Safety, Shandong Center for Disease Control and Prevention, Jinan 250014, China  
Liu Jianwei Institute for Radiological Protection and Safety, Shandong Center for Disease Control and Prevention, Jinan 250014, China  
Liu Haikuan Institute of Radiation Medicine, Fudan University, Shanghai 200032, China liuhk@fudan.edu.cn 
Hits: 4150
Download times: 1966
Abstract::
      Objective To study the methods to estimate absorbed dose to eye lens of infants from CT scanning using different protocols and try to find a practical quick way for estimating absorbed dose to eye lens. Methods By scanning one-year old anthropomorphic phantom using 7 kinds of different protocols, all TLDs were measured for final estimation of absorbed dose to eye lens using two different dose conversion methods. Meanwhile, linear regression equation was established between absorbed dose to eye lens and CTDI. Results Absorbed doses to eye lens of infants from children CT scanning using 7 kinds of different protocols were (9.96±0.69) mGy in head axis, (7.01±0.42) mGy in head helical, (12.60±0.97) mGy in sinus, (12.97±0.42) mGy in inner ear, (0.63±0.03) mGy in neck soft tissue, (8.89±0.44) mGy in cervical vertebra, and (0.34±0.01) mGy in chest,respectively. There were statistically significant difference in doses among different groups (F=846.826,P<0.05). For different scanning locations, there was linear relation between absorbed dose to eye lens and CTDI(r=0.986-0.999, P<0.05). Conclusions Absorbed dose to eye lens of infants from children CT scaning with single dose may not be above threshold dose. In addition, absorbed dose to eye lens can be estimated quickly by linear regression equation between absorbed dose to eye lens and CTDI.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12029849  On-line:0

v
Scan QR Code
&et=08EE5A5866ACE440B2B1C09475B00F966ED829307BF1E6FD51953A45F634CF9794B06689D2038EEE3E1432500C061CD7FD60FEAC51BAF765CB34F476181E6F6A8B38E462EA3B6ABC601B24B4D28BC6700991ED4AE2A4A54163054F4D11260AF433395114CFDC0CDC80FC0BC0E5CFFB92432957350B5D67CC5039EB95DF7CF3B88811EDDA40D2B5CA1A24FA4ED565A847D94C644D7F3AE975&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=B6351343F4791CA3&aid=C287CD5914DB95B014EDB64BDD83C13F&vid=&iid=E158A972A605785F&sid=E42CAFB11D4BE21A&eid=119B6C0AA09DE6B9&fileno=20190410&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="B6351343F4791CA3"; var my_aid="C287CD5914DB95B014EDB64BDD83C13F";