Xu Hui,Yue Baorong,Wei Kedao,et al.Investigation and analysis on the dose of the subjects in CT scanning in China[J].Chinese Journal of Radiological Medicine and Protection,2019,39(3):213-217
Investigation and analysis on the dose of the subjects in CT scanning in China
Received:April 09, 2018  
DOI:10.3760/cma.j.issn.0254-5098.2019.03.010
KeyWords:Computed tomography  CT dose index  Dose length product  Diagnostic reference level
FundProject:卫生行业科研专项(201002009)
Author NameAffiliationE-mail
Xu Hui Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Yue Baorong Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China bryue@163.com 
Wei Kedao Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Wang Jianchao Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Fan Yaohua Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Hits: 5054
Download times: 2101
Abstract::
      Objective To investigate the CT scanning parameters in hospitals at different levels in 15 provinces and cities in China and the doses to patients undergoing CT examination, in order to provide the basis for establishing CT diagnostic reference level suitable for our country. Methods As required in the implementation program for Technical Study on Medical Radiation Hazard Assessment and Control, the information on the patients examined and the CT scanning parameters in clinical practices were investigated. The CT dose index (CTDI100, CTDIW, CTDIVOL) of CT scanner was measured by using the CT ionization chamber. The dose length product (DLP) was calculated on the basis of the scan length of the patients examined. Results A total of 6 524 CT scanning procedures and 483 different types of CT scanning equipment were surveyed in 166 hospitals in 15 provinces in China. For head, chest, abdomen, and lumbar vertebrae, the average weighted CTDIW were 43,15, 19 and 25 mGy, respectively, and the third quartile of CTDIW were 50, 19, 23 and 32 mGy, respectively. The average DLP were 540, 397, 503 and 376 mGy·cm, respectively. The third quartile of DLP were 659, 525, 632 and 479 mGy·cm respectively. Conclusions Through this survey, the doses to CT-examined patients in some provinces were basicly ascertained. The third quartile of doses to four body parts of the examined patients are different from the diagnostic reference level given in publications in other countries and regions. It is important to establish the CT diagnostic reference level suitable for our CT-examined patients according to Chinese national physical characteristics and therefor to promote the optimization of medical radiation protection in CT examination.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12029997  On-line:0

v
Scan QR Code
&et=9A2B3ECE8B55D590AE99BF1AE9328FD22387E875AEADCACCF96947009269EC1CA4D0017E4207FADF20A549A51D787CB84B308A8DF87714DDB80D2CCCFDE87AD8FFE38B0265E95A8B78298FF2A738080C&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=B6351343F4791CA3&aid=C287CD5914DB95B09AF71D79358C6E0F&vid=&iid=38B194292C032A66&sid=527AEE9F3446633A&eid=F9F74EC1AA08A7B9&fileno=20190310&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="B6351343F4791CA3"; var my_aid="C287CD5914DB95B09AF71D79358C6E0F";