Liu Ran,Ma Qiao,Liu Deming,et al.Verification of positioning accuracy of MLC leaves of 7 medical electronic linear accelerators in IMRT in Sichuan province[J].Chinese Journal of Radiological Medicine and Protection,2019,39(2):141-144
Verification of positioning accuracy of MLC leaves of 7 medical electronic linear accelerators in IMRT in Sichuan province
Received:April 24, 2018  
DOI:10.3760/cma.j.issn.0254-5098.2019.02.011
KeyWords:Intensity modulated radiation therapy  Radiochomic film  MLC leaves  Positioning accuracy
FundProject:IAEA资助项目(17821/CRP);四川省科技支撑计划项目(2012SZ0179);四川省卫生与计划生育委员会科研课题(16PJ398)
Author NameAffiliationE-mail
Liu Ran Institute of Occupational Health and Radiological Health, Sichuan Center for Disease Control and Prevention, Chengdu 610041, China  
Ma Qiao Institute of Occupational Health and Radiological Health, Sichuan Center for Disease Control and Prevention, Chengdu 610041, China  
Liu Deming Institute of Occupational Health and Radiological Health, Sichuan Center for Disease Control and Prevention, Chengdu 610041, China  
Liao Xiangdong Institute of Occupational Health and Radiological Health, Sichuan Center for Disease Control and Prevention, Chengdu 610041, China xiangdong987119@163.com 
Hits: 4484
Download times: 1874
Abstract::
      Objective To verify the accuracy of multi-leave collimators (MLC) leaves position in intensity modulated radiation therapy (IMRT) using radiochromic films. Methods 7 medical electronic linear accelerators of Varian, Elekta or Siemens design were chosen. 25 cm×25 cm films were put on 30 cm×30 cm×3.0 cm homogeneous solid phantom and covered with a 2.0 cm thick of homogeneous solid phantom. The CT scanned images were transmitted to TPS for plan preparation. A 5 strip picket fence pattern was formed by MLC leaf, each 3 cm long and 0.6 cm wide, with 3.0 cm wide strip separation. At the maximum dose point, the SSD was 100 cm, with 250 MU per strip. After exposure, the films were transmitted to the dosimetry laboratory at IAEA for measurement and calculation. Results For 6 of 7 accelerators chosen, the differences of film-measured and TPS-planned MLC leaf position for every fence were within ±0.5 mm as required by IAEA, and the other one not consistent with the requirements. The difference of film-measured MLC leaf position between each pair and all pair for 7 accelerators were within ±0.5 mm, in line with IAEA's requirements. The differences of film-measured MLC leaf actual width were within 0.75 mm, as required by IAEA, for 6 accelerators and -0.8 mm for the other one, not consistent with the IAEA requirements. The standard deviations of film-measured MLC leaf actual width for all pairs for 6 accelerators were <0.3 mm, in line with IAEA requirements, but 0.4 mm for the other one, not consistent with IAEA requirements. Conclusions It is simple, fast and accurate to use radiochromic films for verification of the accuracy of MLC leaf position in IMRT. Therefore it is advisable to widely use radiochromic films in IMRT clinical practice.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12040576  On-line:0

v
Scan QR Code
&et=BF7E82E5F022C752D74314D26FEFC0EA0414A96E8BCD90FD7CA50F91F3FF3AFCC8298E238DA7C2C05455AF2C6097EE09629B355AA7F4BE768F9BDDE24729CE6EF390D9596A7FAC67327A5CBF26522DA58DC3050A75F5321D5D8694986A95755B731FCA9FDBA39A7D771D5F8D1DB44AD694FD02DCE07711FF960BAB924C746993&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=B6351343F4791CA3&aid=70F0454B7881B23E026EDAEC60D67D49&vid=&iid=0B39A22176CE99FB&sid=A8DE7703CC9E390F&eid=3986B25773CB6C30&fileno=20190211&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="B6351343F4791CA3"; var my_aid="70F0454B7881B23E026EDAEC60D67D49";