Ren Li,Qiu Rui,Wu Zhen,Ma Ruiyao,Li Junli,Li Chunyan,Su Yinping,Zhu Weiguo,Sun Quanfu.Simulation and analysis of CT examination doses to typical patients based on Chinese reference human phantoms[J].Chinese Journal of Radiological Medicine and Protection,2018,38(12):942-948
Simulation and analysis of CT examination doses to typical patients based on Chinese reference human phantoms
Received:July 19, 2018  
DOI:10.3760/cma.j.issn.0254-5098.2018.12.012
KeyWords:CT  Monte Carlo  Radiation dose  Human phantom
FundProject:
Author NameAffiliation
Ren Li Key Laboratory of Particle & Radiation Imaging, Ministry of Education, Department of Engineering Physics, Tsinghua University, Beijing 100084, China 
Qiu Rui Key Laboratory of Particle & Radiation Imaging, Ministry of Education, Department of Engineering Physics, Tsinghua University, Beijing 100084, China 
Wu Zhen Nuctech Company Limited, Beijing 100084, China 
Ma Ruiyao Key Laboratory of Particle & Radiation Imaging, Ministry of Education, Department of Engineering Physics, Tsinghua University, Beijing 100084, China 
Li Junli Key Laboratory of Particle & Radiation Imaging, Ministry of Education, Department of Engineering Physics, Tsinghua University, Beijing 100084, China 
Li Chunyan Nuctech Company Limited, Beijing 100084, China 
Su Yinping Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Zhu Weiguo Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Sun Quanfu Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China 
Hits: 2508
Download times: 1479
Abstract::
      Objective To simulate and analyze the CT examination doses to typical patients based on Chinese reference human phantom, and to develop an online software for assessing the organ absorbed dose and effective dose to the examinees undergoing CT examination in order to provide a tool for rapid assessment of radiation dose from CT examination.Methods Based on the Monte Carlo simulation, a GE LightSpeed 16 CT scanner was modeled. The organ doses to Chinese adult male, female and 1 year-old child male from single axial scan were calculated based on the Chinese reference phantoms and an organ dose database was established. A web-based CT dose assessment software was developed to read the data from the organ dose database according to the CT scanning parameters entered by the user and to calculate the organ absorbed dose and effective dose rapidly.Results A new CT dose assessment software was developed, which can be used to assess organ doses to Chinese adult male, female and 1-year-old male for different CT scanner, voltage, current, collimator width and other CT parameters. The organ doses calculated in this work were compared with simulation data in two relevant studies. Dose differences for the organs completely included in the scan range were small, with relative differences less than 15% and 25%, respectively.Conclusions The CT dose assessment software provides a powerful tool for the simple and rapid assessment of radiation dose to the examinees undergoing CT examinations.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9058841  On-line:0

v
Scan QR Code
&et=DCB591F1B485B60278AD61206C27C22D5B35099181396841AFE572AF89BE2FD42CE8198EFC801A62B1C301B3453C015C696CF835363B0D7C1128B32A70C7C4768F0729C2F16B8016FE84CC79CA872EEA92B2A65F6B62BBA24DFC68E567F2E8F6C1073099920533E199E3D09099DAEC70&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=BC4A7A631EF4C49932E940F99039DE4B&vid=&iid=59906B3B2830C2C5&sid=72D13CE6C162BDF4&eid=29C36F017AD88B64&fileno=20181212&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="BC4A7A631EF4C49932E940F99039DE4B";