Zhao Hongfu,Zhao Zhipeng,Han Dongmei,et al.A method for determination of dwell position offset of ring applicator using GafchromicTM EBT3 film and ImageJ software[J].Chinese Journal of Radiological Medicine and Protection,2018,38(11):865-869
A method for determination of dwell position offset of ring applicator using GafchromicTM EBT3 film and ImageJ software
Received:April 10, 2018  
DOI:10.3760/cma.j.issn.0254-5098.2018.11.013
KeyWords:Brachytherapy  Applicator reconstruction  GafchromicTM EBT3  ImageJ
FundProject:国家自然科学基金项目(31600679)
Author NameAffiliationE-mail
Zhao Hongfu Department of Radiation Oncology, China-Japan Union Hospital of Jilin University, Changchun 130033, China  
Zhao Zhipeng Department of Radiation Oncology, China-Japan Union Hospital of Jilin University, Changchun 130033, China  
Han Dongmei Department of Radiation Oncology, China-Japan Union Hospital of Jilin University, Changchun 130033, China  
Cheng Guanghui Department of Radiation Oncology, China-Japan Union Hospital of Jilin University, Changchun 130033, China chengghcjuh@sina.com 
Hits: 3782
Download times: 1879
Abstract::
      Objective To report a simple method for determination of dwell position offset of ring applicator using GafchromicTM EBT3 film and ImageJ software. Methods In the dummy plan, this study reconstructed the ring applicator point by point manually according to the MR line marker, the most distal reconstruction point was at the end of marker, set the offset to zero. The first, eleventh, and twenty-third dwell positions were activated with the step length of 2.5 mm. A setup image was designed to immobilize the ring applicator. The setup image was composed of two perpendicular thin lines and a circular with a diameter of 16 mm. GafchromicTM EBT3 film was fixed behind the setup image by plastic tapes. After irradiation, a ruler was put on the bottom of the irradiated film as a plotting scale. Photos were taken with the help of a vertical camera using a tripod. The file was imported into ImageJ software and the coordinate values were obtained by semi-automatic recognition of the center of exposure area. The test was repeated five times at one week interval to evaluate the uncertainty. For every time, each dwell position was measured five times. Results The angle between the radius of the first dwell position and the x-axis was 43° in the dummy plan, and (46.8±0.9)° for actual measurement. The offset was 0.9 mm for the ring applicator combined with the MR according to calculation. The distances between the first,the eleventh and the twenty-third dwell positions and the center of the ring were (12.9±0.1) mm, (12.5±0.1) mm, (12.4±0.1) mm, respectively. Conclusions Using GafchromicTM EBT3 film and the ImageJ software is a simplified method to validate the dwell position offset of the ring applicator.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12030146  On-line:0

v
Scan QR Code
&et=2D8A5CE31A4A632BCB1C353B8B55997DD2157B3A143F76AB01296D2C4D71CD497CA2BD24EB2115A0C7F0CD4E929DCC60A4D647A7283D1D7A586E3E526AE74157C83A19DB7ADAE0EE366B5E028D99A90FF854D5D7A19EC33529B4D78E9F254CD6105D73744F0567FBFE57507F50AAA7AF0F195E0B35ABDC8F239672DF9903036C4A69AFB7101BEFCC&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=BC4A7A631EF4C4997D7E2BECBA335E73&vid=&iid=708DD6B15D2464E8&sid=461E94ABCF58C63F&eid=F26986CDF689DBC4&fileno=20181113&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="BC4A7A631EF4C4997D7E2BECBA335E73";