Zhang Dandan,Zhang Qun,Chen Bin,et al.Comparison of digital radiography's entrance surface doses computed from DICOM file or directly from TLDs[J].Chinese Journal of Radiological Medicine and Protection,2018,38(10):787-791
Comparison of digital radiography's entrance surface doses computed from DICOM file or directly from TLDs
Received:February 24, 2018  
DOI:10.3760/cma.j.issn.0254-5098.2018.10.014
KeyWords:Medical irradiation  Entrance surface dose  Digital radiography  DICOM file
FundProject:
Author NameAffiliation
Zhang Dandan Department of Environmental and Occupational Health Promotion, Ningbo Municipal Center for Disease Control & Prevention, Ningbo 315010, China 
Zhang Qun Department of Environmental and Occupational Health Promotion, Ningbo Municipal Center for Disease Control & Prevention, Ningbo 315010, China 
Chen Bin Department of Radiology, Ningbo Second Hospital, Ningbo 315010, China 
Wang Aihong Department of Environmental and Occupational Health Promotion, Ningbo Municipal Center for Disease Control & Prevention, Ningbo 315010, China 
Hu Bibo Department of Radiology, Ningbo Second Hospital, Ningbo 315010, China 
Qu Lingyan Department of Environmental and Occupational Health Promotion, Ningbo Municipal Center for Disease Control & Prevention, Ningbo 315010, China 
Hits: 4210
Download times: 2129
Abstract::
      Objective To make comparison of consistency between the ESD's information-calculated values and TLD readout values based on PHILIPS DigitalDiagnost so as to provide reference for realizing information management of medical exposure radiation dose during DR diagnosis and treatment. Methods TLDs were used to obtain the ESD values on chest, abdomen, vertebral body, pelvis and limbs. The exposure parameters were collected from the DICOM files, and the information value of ESD were calculated through the dose area product (DAP) method and the exposure parameter method. The consistency between information value and TLD readout value was compared by intraclass correlation coefficient and the Bland-Altman method. Results The ESD values obtained from DAP method, exposure parameter method and TLD method are, respectively, (0.730±0.870) mGy, (0.720±0.863) mGy and (0.770±0.948)mGy. ICC was 0.988 for ESDDAP-TLD and 0.982 for ESDParameters-TLD, respectively. The Bland-Altman map showed that the D-value of ESD obtained from DAP-TLD and Parameters-TLD beyond the 95% consistency interval was 6.7%. Conclusions The PHILIPS DR daily diagnosis and treatment process is based on fixed-mode programmed diagnosis and treatment. Under the premise of qualified equipment quality control, the information calculated value based on DICOM files has a good consistency with TLD readout value.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12031079  On-line:0

v
Scan QR Code
&et=8C10CEA789A144B8281AEADD8B83F7381CB464357FB514DAA08AF7B103994BA90A39B62DAD9E14F8365833AB516E9B95D69EE316DF941FA47785D5443AA0882309888C22113EA0E6C6A6D03CB35609C542900EB68A8E54B0894A3D771340AA9C7CD03FF43F5C0DC97D707460A14967DD&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=BC4A7A631EF4C499BC6B5F48B6142D9F&vid=&iid=F3090AE9B60B7ED1&sid=872F6C582A30BA57&eid=4198A31627C9B2A6&fileno=20181014&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="BC4A7A631EF4C499BC6B5F48B6142D9F";