Liu Dandan,Cui Ying,Zhao Bo,Zhang Yongxian,Niu Yantao.The phantom study of the effect of tube voltage on radiation dose and image quality of CT with automatic tube current modulation mode[J].Chinese Journal of Radiological Medicine and Protection,2018,38(9):710-714
The phantom study of the effect of tube voltage on radiation dose and image quality of CT with automatic tube current modulation mode
Received:January 24, 2018  
DOI:10.3760/cma.j.issn.0254-5098.2018.09.014
KeyWords:Tomography, X-ray computed  Radiation dose  Tube voltage
FundProject:北京市卫生系统高层次卫生技术人才培养计划(20143019);北京市医院管理局临床医学发展专项经费资助("扬帆"计划眼耳鼻喉影像重点医学专业)(ZYLX201704)
Author NameAffiliationE-mail
Liu Dandan Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Cui Ying Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Zhao Bo Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Zhang Yongxian Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China  
Niu Yantao Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China ytniu163@163.com 
Hits: 3584
Download times: 1542
Abstract::
      Objective To investigate the effect of the change of tube voltage on radiation dose and image quality in head-neck and chest scanning under automatic tube current modulation(ATCM). Methods CT scanning was performed on the head-neck and chest phantom with ATCM and automatic tube voltage(CARE kV). The tube voltage was manually selected at 70, 80, 100, 120 and 140 kV separately, and a routine CT scanning of the head-neck and chest with ATCM was performed. The scout was scanned for 3 times and a spiral scanning was performed once at each of tube voltage. The regions of interest(ROIs) were selected in the slices of orbital center and C5 upper edge level for the head-neck phantom, in the slices of apical and tracheal bifurcation level for the chest phantom. The contrast to noise ratios (CNRs) were measured and recorded. The organ dose of eye lens and mammary are measured with thermoluminescent dosimeters (TLDs) for every scanning (the average of 3 measurements). The cumulative dose value of the scout and spiral scanning was calculated. The volume CT dose index (CTDIvol) of each scan was recorded, and the cumulative value of CTDIvol was calculated. Finally, the optimized tube voltage was obtained by calculating the FOM (figure of merit). Results With ATCM and CARE kV, 120 kV and 108 mAs were chose automatically by system for head-neck phantom, 80 kV and 167 mAs for chest phantom. With ATCM, the radiation dose of eye lens and CTDIvol were minimal with manually selected 70 kV (0.779 and 4.070 mGy respectively),and maximaum with manually selected 140 kV (2.571 and 25.670 mGy). The radiation dose of the mammary gland and CTDIvol were minimal with manually selected 70 kV (0.698 and 0.900 mGy), and maximal with manually selected 140 kV (3.452 and 7.400 mGy). The CNR values of orbital center and C5 upper edge level were 51.30-118.36 and 80.78-173.12 respectively. The CNR values of the apical and tracheal bifurcation level were 50.15-129.58 and 49.63-115.40, respectively. The optimal FOM was appeared at orbital center slice with 80 kV, at C5 upper edge level slice with 120 kV and at both the apical and tracheal bifurcation level with 70 kV. Optimum tube voltage for head-neck phantom:manual 100 kV at orbital level, CARE kV mode(120 kV) at neck level. Optimal tube voltage for chest phantom:manual 100 kV. Conclusions The selection of tube voltage is responsible for the radiation dose and image quality of CT scanning. For conventional CT scan, manual 100 kV is suitable for orbital scanning, automatic 120 kV is suitable for neck scanning, manual 100 kV is suitable for chest scanning.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9073951  On-line:0

v
Scan QR Code
&et=9BCA4CC1B42B911BC107562AA343A3489F3B9E6CAF84CB9827EFC5E15BC40D494CECE25A64B7E3DDBB6AC656FF25AEEE2B4DF9C169A909AFB30AE5495D866BDBF3404813FE43AC02CBC77519478BB5D963B9AB19140891E11E4A6D7E6A764EEBE0ED99DE5C467BA13E592F23B213B7F8885BDF53951C238728480D91553443737F16096751BE35B7&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=415CD46E06B7C80C2ECDB12E81546407&vid=&iid=9CF7A0430CBB2DFD&sid=2868AE484E7BAD6B&eid=C4B91B15F1F73E7E&fileno=20180914&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="415CD46E06B7C80C2ECDB12E81546407";