Fan Lijuan,Xu Dongsheng,Zhang Jiwang,Liu Junbo,Fu Donghai,Ling Jian,Yu Tielian.The investigation on image quality and radiation dose of low dose wider detector array CT scanning for young children with congenital heart disease[J].Chinese Journal of Radiological Medicine and Protection,2018,38(8):626-630
The investigation on image quality and radiation dose of low dose wider detector array CT scanning for young children with congenital heart disease
Received:March 12, 2018  
DOI:10.3760/cma.j.issn.0254-5098.2018.08.013
KeyWords:Computed tomography  Congenital heart disease  Radiation dosage  Iterative reconstruction
FundProject:
Author NameAffiliationE-mail
Fan Lijuan Department of Radiology, TEDA International Cardiovascular Disease Hospital, Tianjin 300457, China  
Xu Dongsheng Department of Radiology, TEDA International Cardiovascular Disease Hospital, Tianjin 300457, China  
Zhang Jiwang Department of Radiology, TEDA International Cardiovascular Disease Hospital, Tianjin 300457, China  
Liu Junbo Department of Radiology, TEDA International Cardiovascular Disease Hospital, Tianjin 300457, China  
Fu Donghai Department of Radiology, TEDA International Cardiovascular Disease Hospital, Tianjin 300457, China  
Ling Jian Department of Radiology, TEDA International Cardiovascular Disease Hospital, Tianjin 300457, China  
Yu Tielian Department of Radiology, General Hospital of Tianjin Medical University, Tianjin 300052, China tjzyyytl@163.com 
Hits: 2718
Download times: 1339
Abstract::
      Objective To investigate the image quality and radiation dose of the wider detector array CT scanner with low dose scanning mode in young children with congenital heart disease.Methods Totally 100 consecutive pediatric patients younger than 3 years with congenital heart disease were enrolled. They were divided into two groups. The low dose group with fifty patients underwent axial CT scanning with ECG gating, and the control group with fifty patients were scanned with volume helical shuttle (VHS) technique. CT number and noise of two groups images at the level of ascending aorta, main pulmonary artery, left ventricle, descending aorta and adjacent muscle were measured, and the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. The double blind method was used to evaluate subjective image quality of the level of intra-cardiac, extra-cardiac and coronary artery. Effective dose was also calculated for both groups.Results No significant difference was found in the CT number, image noise, SNR, and CNR between the two groups in the same anatomic regions(P>0.05). No significant difference was found in subjective image quality between the two groups for the intra-cardiac and extra-cardiac structure(P>0.05). The subjective image quality of coronary artery was significant higher in low dose group than the control group(4.10±0.90 vs.2.88±0.82, Z=-5.818, P<0.05). Effective dose was (0.57±0.30)mSv in group A and (2.39±1.15)mSv in group B with dose savings of 76%(t=-11.642, P<0.05).Conclusions The wider detector array CT scanner with low dose scanning mode can improve image quality with lower radiation dose.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9057854  On-line:0

v
Scan QR Code
&et=640A177CCC182B56AF4FC73398B91DF13E5A4B732D2DD503ADBD8236626D417CAFB6B98D2910D59FB0BCFB8CD1C830DB26D1ADBB37735BC4AE295126B338D2B77516F8E0DB2135D0C877A82D4AF83854961B6D65ED70B8A11C5D257E84EA443B65322BA6C0F717E3AEFF3CAB37DB73F23CE3872E88857B3B76D1CDB7D65D9CD9B8652296892FC15331A0AED065620D17CF86FC751B081751&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=415CD46E06B7C80C5384AD3B33CADB96&vid=&iid=5D311CA918CA9A03&sid=C824C8F9F54AE9B9&eid=039DCCB9394D9766&fileno=20180813&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="415CD46E06B7C80C5384AD3B33CADB96";