Verification of absorbed dose and single-field dosimetry of MLC-shaped IMRT photon beams in Hubei province
Received:August 17, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2018.02.010
KeyWords:Intensity modulated radiotherapy  Thermoluminescent dosimeter  Film  Absorbed dose  Two-dimensional dose distribution
FundProject:IAEA基金资助项目(17821/CRP)
Author NameAffiliationE-mail
Zhou Wenshan Hubei Provincial Center for Disease Control and Prevention, Wuhan 430079, China  
Sun Gangtao Hubei Provincial Center for Disease Control and Prevention, Wuhan 430079, China  
Ye Song Hubei Provincial Center for Disease Control and Prevention, Wuhan 430079, China  
Luo Suming Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China sumingluo@163.com 
Hits: 3539
Download times: 1895
Abstract::
      Objective To verify the absorbed dose and two-dimensional dosimetry delivered by MLC-shaped IMRT photon beams using thermoluminescent dosimeters (TLDs) and radiochromic films. Methods Different types of medical linear accelerators at 7 third-grade first-class hospitals in Hubei were selected for this audit study. Besides TLDs and films, a small polystyrene phantom measuring 15 cm×15 cm×15 cm, supplied by the International Atomic Energy Agency (IAEA), was also used. The TPS dose calculation were performed with 6 MV X-ray beam and by delivering 6 Gy dose to the TLDs or films at 10 cm depth for 5 cm×5 cm field size at 90 cm SSD. After irradiation, comparations were conducted of dose derivations between TPS-calculated values and TLD-or film-measured values. The two-dimensional dosimetry audit was carried out by using 30 cm×30 cm homogeneous solid phantoms in which a 25 cm×25 cm film was located at 5 cm depth. A single-filed profile generated by TPS was delivered to the film, and the film was measured. And then, two-dimensional dose distribution was analyzed using the film analysis system. Results The results suggested that the relative deviations of TLD and film to TPS dose was -8.5% and -1.9% for accelerator 1, 5.4% and 0.5% for accelerator 7, respectively, whereas those for other accelerators were within ±5%. The passing rates for two-dimensional dose distribution at all accelerators were all higher than 90%. Conclusions TLD and film quality audit for IMRT can be applied for a wide range of application in view of its simplicity, reliability and postal convenience.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9074778  On-line:0

v
Scan QR Code
&et=BF7E82E5F022C752D74314D26FEFC0EA9B0FD975A45D275CAF807E64ED21FFF107F0F5514F28C64350E4F62866BADA5FD579D9D37993F8577C4BB76260E4AFA6A4FD79163F0704B6C6DADB589244DC412A6535DD3CE4D43A49C35B99904A693A2FCCDA119839515FE030859168C52B43&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=295B27E71BBC5E126931B6FE58EFCEEF&vid=&iid=0B39A22176CE99FB&sid=28F8B56DB6BEE30E&eid=76B5E24D6EC46B4B&fileno=20180210&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="295B27E71BBC5E126931B6FE58EFCEEF";