Wang Jin,Yang Chunyong,Zhou Yuanyuan,Yu Ningle.Dosimetric verification on intensity modulated radiotherapy in Jiangsu province[J].Chinese Journal of Radiological Medicine and Protection,2018,38(2):121-124
Dosimetric verification on intensity modulated radiotherapy in Jiangsu province
Received:July 26, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2018.02.008
KeyWords:Intensity modulated radiotherapy  Dose verification  Thermoluminescent dosimeter  Film
FundProject:IAEA资助项目(17821/CRP);江苏省流行病学重点学科(ZDXK A2016008);江苏省医学创新团队(CXTD A2017029)
Author NameAffiliationE-mail
Wang Jin Jiangsu Provincial Center of Disease Control and Prevention, Nanjing 210009, China  
Yang Chunyong Jiangsu Provincial Center of Disease Control and Prevention, Nanjing 210009, China  
Zhou Yuanyuan Jiangsu Provincial Center of Disease Control and Prevention, Nanjing 210009, China  
Yu Ningle Jiangsu Provincial Center of Disease Control and Prevention, Nanjing 210009, China yuningle@163.com 
Hits: 3403
Download times: 1645
Abstract::
      Objective To explore the dosimetric verification method for absorbed dose and two-dimensional dose distribution generated through multi-leaf collimator in radiotherapy. Methods A total of 8 medical accelerators were selected in Jiangsu province. Under 6 MV X-rays with a 5 cm×5 cm field, the thermoluminescent dosimeter (TLD) and EBT3 radiochromic film (RCF) were used to audit the values calculated by treatment planning system (TPS). RCF was used to verify two-dimensional dose distribution under the 6 MV MLC beams for intensity modulated radiation therapy (IMRT). Results The relative dose deviation was from -1.4% to 6.8% for TLD and from -1.8% to 7.8% for RCF, with 7 accelerators meeting with the IAEA deviation requirement of ±5%. The passing rates of two-dimensional dose distribution were between 73.7% and 97.0%, with 7 accelerators greater than 90%. Conclusions TLD and RCF are feasible and suitable for large-scale verification of IMRT TPS absorbed dose and two-dimensional dose distribution.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9072525  On-line:0

v
Scan QR Code
&et=8400690B225D0A44A3980D25683E7AFD04FA9AD1630B752609BC7EFAEB85DFB3212CAE930EF8B080D378AEB79AD75CA52A4320FF0F49BEF931348A63F71097F6F3C43B60DA9D16DEC516C0C3AE00F677&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=295B27E71BBC5E1268D1DBFCA30AB33E&vid=&iid=0B39A22176CE99FB&sid=CDEBD1ACE0A4C1C1&eid=2F56B21F91C9B05B&fileno=20180208&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="295B27E71BBC5E1268D1DBFCA30AB33E";