Zhang Zhen,Lian Dexing,Hou Changsong,Zhang Qingzhao,Xu Hui,Zhu Weiguo,Liang Jing.Validation and discussion of the time response correction formula for radiation detectors[J].Chinese Journal of Radiological Medicine and Protection,2018,38(1):48-51
Validation and discussion of the time response correction formula for radiation detectors
Received:September 27, 2017  
DOI:10.3760/cma.j.issn.0254-5098.2018.01.010
KeyWords:Time response  Ambient dose equivalent rate  Correction formula
FundProject:
Author NameAffiliationE-mail
Zhang Zhen Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Lian Dexing Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Hou Changsong Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Zhang Qingzhao Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Xu Hui Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Zhu Weiguo Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China  
Liang Jing Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China liangjing_cdc@163.com 
Hits: 2941
Download times: 2908
Abstract::
      Objective To validate and discuss the time response correction formula for four types of dosimeters (6150AD6+6150AD-b, FH40G+FHZ672E-10, 451P ionization chamber and AT1123). Methods The ambient dose equivalent rates shown by survey meters were recorded separately when X-ray emission time was 500, 200, 100 and 50 ms.The corrected values were obtained by the formula of circuit having a capacitance C and asistance R in series. Results Therewas no correlation between the value measured by AT1123 dosimeter and the time of irradiation.The values by other three kinds of dosimeters obviously varied with the time of irradiation. Conclusions It is not required to make the time response correction for the measured value of AT1123 dosemeter, whereas the values measured by the other three dosimeters could be corrected by the time response correction formula.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9066229  On-line:0

v
Scan QR Code
&et=315CD8DD583D2E3B0A982FD0FC468FF5D0095AB1ACA641B2BD93F71B44E7E59D7FA229ED55A963F1799853F9FC1BF5A64676BADA8BB6930AC4A814954C6C0E6C01E798B62D9CBD6BC22A2F1AC5293374D306E94F8DFAA53B6FBB3F427D19F19B&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=EA357AD73C8E13BC&aid=7B1F7255187612B0FD12141E52F1799C&vid=&iid=CA4FD0336C81A37A&sid=B6DA1AC076E37400&eid=987EDA49D8A7A635&fileno=20180110&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="EA357AD73C8E13BC"; var my_aid="7B1F7255187612B0FD12141E52F1799C";