Zhang Zhen,Lian Dexing,Hou Changsong,et al.Validation and discussion of the time response correction formula for radiation detectors[J].Chinese Journal of Radiological Medicine and Protection,2018,38(1):48-51 |
Validation and discussion of the time response correction formula for radiation detectors |
Received:September 27, 2017 |
DOI:10.3760/cma.j.issn.0254-5098.2018.01.010 |
KeyWords:Time response Ambient dose equivalent rate Correction formula |
FundProject: |
Author Name | Affiliation | E-mail | Zhang Zhen | Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | | Lian Dexing | Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | | Hou Changsong | Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | | Zhang Qingzhao | Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | | Xu Hui | Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | | Zhu Weiguo | Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | | Liang Jing | Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | liangjing_cdc@163.com |
|
Hits: 3735 |
Download times: 3411 |
Abstract:: |
Objective To validate and discuss the time response correction formula for four types of dosimeters (6150AD6+6150AD-b, FH40G+FHZ672E-10, 451P ionization chamber and AT1123). Methods The ambient dose equivalent rates shown by survey meters were recorded separately when X-ray emission time was 500, 200, 100 and 50 ms.The corrected values were obtained by the formula of circuit having a capacitance C and asistance R in series. Results Therewas no correlation between the value measured by AT1123 dosimeter and the time of irradiation.The values by other three kinds of dosimeters obviously varied with the time of irradiation. Conclusions It is not required to make the time response correction for the measured value of AT1123 dosemeter, whereas the values measured by the other three dosimeters could be corrected by the time response correction formula. |
HTML View Full Text View/Add Comment Download reader |
Close |
|
|
|